This is an intermittent problem with autotuning to frequencies for which the Coarse knob is set to X.
Autotune just does not work and does not show any errors. This means we are not able to observe
nuclei like 31P or 11B. However, we can tune manually to these frequiencies, and autotune works for
the other low-band nuclei. Pre-tune also fails (because it starts with tuning to 31P) showing the following
error: "LF coarse dial not set".

this is a bit strange!

1. Have restarted the console?
2. All thecables from the stepping motors are tightly connected to the probe?
3. Have you tried an other probe - and the same troubles coming up?

Maybe something in the probe is broken. These are quite long lines out of glassfibre material. Or the
condensator on the top ofthis bar is broken. All can happen - this is the mechanics inside.

Keep trying and finding the fault.

Yours sincerely,

Ulrich Haunz



Has any ECS user observed such issue? If yes, how did you fix it?
Thank you,
Piotr