Re: [AMMRL] Gradient calibration frequency

From: Charles Fry via groups.io <cfry03=esf.edu_at_groups.io>
Date: Tue, 1 Sep 2026 15:03:01 +0000

Alex,

The Bruker Diffusion manual (available in Topspin) gives a good description
on how to calibrate the gradients by using standard samples. For liquids see
section 2.10 to begin with. I’ve done this a few times and it works well.
I check my VT calibrations close to ambient temps at the same time.

I agree that the calibration doesn't seem to change if everything is stable,
but doing this at least once for any probe doing diffusion work should be
done, as the Bruker default can be off. Josh Kurutz has a nice description
of dosy restore on the U Chicago site (at the top of the blog), used when
the gradient calibration is changed.

Charlie


> From: main_at_ammrl.groups.io On Behalf Of Alex Kitaygo rodskiy via groups.io
> Sent: Sunday, August 30, 2026 5:37 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Gradient calibration frequency

Hi all,

I’d be interested to hear how other facilities handle calibration of
Z-gradients for PFG diffusion experiments.

How often do you routinely check or recalibrate the gradient strength on your
liquid probes (we have NEO instruments with Prodigy and Smart probes)? Is
this an annual PM item, done per project, or only after major system/probe
servicing?

Have you observed noticeable physical degradation in gradient coil performance
or amplifier output over time without explicit hardware failure?

Any insights or standard operating procedures would be greatly appreciated.

Thanks,

Alex


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3731): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3731__;!!PvDODwlR4mBZyAb0!RN8EIJ_b8O-jCt7lHUNY-wSsVQ-4ha=
6RQ7gVyPqog5eO2FjuKMcc8WSEeGWZOPnMU5Mhfltfy6sYdM3Q6ChUp4tpig$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121012683=
/7559972__;!!PvDODwlR4mBZyAb0!RN8EIJ_b8O-jCt7lHUNY-wSsVQ-4ha6RQ7gVyPqog5eO2=
FjuKMcc8WSEeGWZOPnMU5Mhfltfy6sYdM3Q6Ch35Cuvtw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Sep 01 2026 - 08:08:45 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 02 2026 - 02:30:34 MST