Hi Alex,
In my (limited) experience with high-resolution solution state probes on GAB/2
with AV III / AV III HD, and GAB/2 (AV Neo), I've not seen any changes over
time, even after probes have gone back to Bruker for repairs on the RF side.
Having said that, if something breaks or deteriorates over time you won't
know about it until you actually verify those gradient strengths - it's
not like running (say) ledbpgp2s takes very long.
Hopefully you'll get more anecdata from other facility managers on this, but
I think the only way you're going to be able to convince yourself for a given
system is to make those measurements yourself over time.
For the most consistent measurements over multiple years you need to use
flame-sealed samples and preferably the same physical samples as well. You
want to verify your sample temperatures first (e.g. 99.8% MeOH-d4), and only
then calibrate your gradients using a sample with well-defined physical
properties e.g. doped water.
Thanks
Geoff
> From: main_at_ammrl.groups.io On Behalf Of Alex Kitaygorodskiy via groups.io
> Sent: 30 August 2026 23:37
> To: main_at_ammrl.groups.io
> Subject: [External] [AMMRL] Gradient calibration frequency
Hi all,
I'd be interested to hear how other facilities handle calibration of Z-gradients
for PFG diffusion experiments.
How often do you routinely check or recalibrate the gradient strength on your
liquid probes (we have NEO instruments with Prodigy and Smart probes)? Is
this an annual PM item, done per project, or only after major system/probe
servicing?
Have you observed noticeable physical degradation in gradient coil performance
or amplifier output over time without explicit hardware failure?
Any insights or standard operating procedures would be greatly appreciated.
Thanks,
Alex
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3730): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3730__;!!PvDODwlR4mBZyAb0!SDSQYpnAHtplMJ7CTf49KjBlkte8S_=
6sX8VkKJjP2d5-koy_if2bQYY4drwZrYJPfBL9Bf-CGtjHiaNMeP46JpA63eqf$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121012683=
/7559972__;!!PvDODwlR4mBZyAb0!SDSQYpnAHtplMJ7CTf49KjBlkte8S_6sX8VkKJjP2d5-k=
oy_if2bQYY4drwZrYJPfBL9Bf-CGtjHiaNMeP46JsZuKjCU$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Sep 01 2026 - 07:12:03 MST