Re: [AMMRL] continuing interrupted 2D, nD experiments

From: Ulrike Werner-Zwanziger via groups.io <Ulli.Zwanziger=Dal.ca_at_groups.io>
Date: Wed, 17 Sep 2025 10:32:21 -0300

Hello Charlie,

I have extended 2-dimensional data in the past, though not in a situation
you are in. I did it when I performed T1-experiments and found that I
needed more points. The strategy is that you can create FID's from the ser
file and then combine those again into a ser file.

In detail:
- Copy the 2d-data to an experiment number where sufficient experiment
numbers directly following it are not used. The au program "splitser" then
creates the corresponding number of FID's.
- Do the same for the follow up experiment that contains the rest of the
data
- Copy the latter into experiments that directly follow the first ones.
- "fidtoser" then will create a new 2d data set. Its au program header
describes how to choose the parameters and how to do it.

In my situation it was guaranteed that the experimental parameters like
phases, excitation powers, etc were the same, as I performed the
experiments right after the first ones. I am not sure that these conditions
are given, after a power outage.

Best of luck and happy resonating,

Ulli


On Wed, Sep 17, 2025 at 6:50 AM Charles Fry via groups.io wrote:

> Hi all,
>
> Was asked by a user about this recently. My perhaps outdated experience
> tells me that a 2D that is interrupted by a power outage or some other
> event cannot be continued. But that's maybe just the simple answer. Figured
> if it can be done, someone here would know how.
>
> An example: suppose TD1 is set to 400 but the experiment is interrupted
> at 171. Can we set the first delay and phase cycling (and gradients and
> ...?) properly and continue in obtaining 172 to 400, and then add that on to
> the dataset? Clearly would be easier for some experiments than others. But
> for now, I'm interested in whether there's some way to get topspin so it
> can do this for even the simplest of 2Ds: i.e., acquire the 172 to 400
> data. A separate issue, but necessary, is then how to tag that onto the 1
> to 171 data.
>
> Charlie
>

--
Dr. Ulrike Werner-Zwanziger
Adjunct Professor
Department of Chemistry, NMR-3,
Dalhousie University
6245 Castine Way (!!!new address!!!)
PO BOX 15000
Halifax, NS B3H 4R2

e-mail: Ulli.Zwanziger_at_Dal.ca
Tel. 1 902 494 8085
FAX: 1 902 494 1310


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2818): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2818__;!!PvDODwlR4mBZyAb0!VidlKBOUc52-Xcc6dcFeCRGZOvDpXR=
poTU1QdvgQSlgBvXi-uh0sYUiIi3XLoKDrXacpPAMf33VAzLcOFhw4lhVEKU3-$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115288518=
/7559972__;!!PvDODwlR4mBZyAb0!VidlKBOUc52-Xcc6dcFeCRGZOvDpXRpoTU1QdvgQSlgBv=
Xi-uh0sYUiIi3XLoKDrXacpPAMf33VAzLcOFhw4lj4KHilg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Sep 17 2025 - 06:44:47 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 01 2025 - 14:23:05 MST