Just a minor refinement of Mike's answer — something I learned from Quantum
Design because I also care for a SQUID — instead of a wooden dowel, you can
use a shotgun cleaning rod. It is generally either brass or aluminum, and it
has a plastic eye on the end which works great for threading a kim-wipe through
for cleaning. It also has a T handle on top, which is convenient for holding.
As with the dowel, soak the kim-wipe with your solvent of choice. They turn
out to be a great length for 400-500 MHz magnet, and they are in screwable
sections if you need one for a larger magnet.

--Cathy


Catherine F. M. Clewett, Ph.D.
Asst. Director Magnetic Resonance
Chemistry Instrumentation Center
University of Wisconsin Madison, Department of Chemistry
1101 University Ave Room 2201c, Madison, WI 53706

clewett_at_wisc.edu, (608)262-8196, https://pbcic.chem.wisc.edu



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2813): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2813__;!!PvDODwlR4mBZyAb0!Sd-EdSXP6o6OMFFeTGWauAQIgRhvQR=
f5Lmp-AhLWDfmduxSjubmSlxU4XPZWEBZa-R1RY0sLfFuoyNOuHdiC808yAsNM$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115256205=
/7559972__;!!PvDODwlR4mBZyAb0!Sd-EdSXP6o6OMFFeTGWauAQIgRhvQRf5Lmp-AhLWDfmdu=
xSjubmSlxU4XPZWEBZa-R1RY0sLfFuoyNOuHdiC8-O7MSgT$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Sep 16 2025 - 08:44:46 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 01 2025 - 14:23:05 MST