Happy to report that increasing the air flow via edte worked fine.
For our system, I’m told it can be increased up to 1500. I tried
increasing to 900, and the sample came up fine.
Have a great week,
Jim
--
Jim Baleja
(he/him/his)
Associate Professor
Departments of Medical Education (office: MV525S) and Developmental, Molecu=
lar, & Chemical Biology (lab: MV614)
Tufts University School of Medicine
136 Harrison Avenue
Boston, MA 02111
Phone: 617-636-6872 (goes to voice mail)
https://pubmed.ncbi.nlm.nih.gov/
> From: main_at_ammrl.groups.io on behalf of Baleja, James D via groups.io
> Date: Monday, September 8, 2025 at 11:53 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [External] [AMMRL] Code to temporarily increase air for sample ejection
Our house air service is periodically just a PSI or 2 short and we can't eject
samples. I'm thinking that I may be able to increase the ejection airflow a
tiny amount temporarily. However, I don't have the service code for the BSMS.
Is there a generic code? We're running Topspin 3.6 on our Bruker.
Thanks Jim
-=-=-=-=-=-=-=-=-=-=-=-