Re: [AMMRL] Code to temporarily increase air for sample ejection

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Mon, 8 Sep 2025 16:45:41 +0000 (UTC)

 Hi Jim,

The super secret code is (write this down) 1964.  That will get you into
the service functions of the BSMS.  

Note, there are other super secret codes that are not 1964 so you can't just
use this willy-nilly for everything.

If the eject gas is a little low then another alternative if you're trying
to eject it manually is to crank up the VT gasflow--1500 lpm is frequently
enough to get the sample out.  If you try to do this with a samplecase
though, it will get mad.

Cheers,Mike

On Monday, September 8, 2025 at 09:53:49 AM MDT, Baleja, James D via groups.io wrote:

Our house air service is periodically just a PSI or 2 short and we can't eject
samples.  I'm thinking that I may be able to increase the ejection airflow
a tiny amount temporarily.  However, I don't have the service code for the
BSMS.  Is there a generic code? We're running Topspin 3.6 on our Bruker. 

Thanks 

Jim 


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2775): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2775__;!!PvDODwlR4mBZyAb0!QrWysw3lnNp3f_OQZybofqb46fDFaK=
OqEVGfSJVb6I-KY__a1voim_hTl_CMjGPNOaBx7Zbqj9FXoJkTIGqNpSsT1p4w$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115133221=
/7559972__;!!PvDODwlR4mBZyAb0!QrWysw3lnNp3f_OQZybofqb46fDFaKOqEVGfSJVb6I-KY=
__a1voim_hTl_CMjGPNOaBx7Zbqj9FXoJkTIGqNpfqS5u4X$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Sep 08 2025 - 09:50:27 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 01 2025 - 14:23:05 MST