Hi Brian,

Roughing pump should be fine, no need for a turbo.

The biggest QOL improvement comes from attaching a vacuum gauge. Indispensable
for knowing achievable end point, measuring vacuum progress, and re-testing
for softening. Unfortunately they run upwards of $800. I'd check with coworkers
before buying new; any synthetic lab doing schlenk work would have several in use.

Kind regards,
Luke

Luke Fulton, PhD
CHEM BLDG R003
NMR Core Facility Director
Unit 3060
COR2E & Department of Chemistry
55 N Eagleville Road
University of Connecticut
Storrs, CT 06279

email: fko24003_at_uconn.edu
(alias): luke.fulton_at_uconn.edu
mobile: (603) 953-5275
Office: (860) 486-4069
________________________________

> From: main_at_ammrl.groups.io on behalf of Brian Antalek via groups.io
> Sent: Wednesday, September 3, 2025 1:46 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] vacuum attachment for He transfer line


I have a He transfer line that has softened and needs to be pumped down. It's
from a 2005 ish Bruker AVII 300 MHz instrument and the vacuum connector piece
cannot be found (see attached photo). Does anyone know where I can find one?
Also, will a rough pump be enough, in general?

Thanks
Brian


Brian Antalek

Bausch + Lomb, Inc.
1400 N. Goodman Street
Rochester, NY 14609
http://www.bausch.com

brian.antalek_at_bausch.com<mailto:brian.antalek_at_bausch.com>
(585) 752-8452 mobile
(585) 666-1127 Teams



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2762): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2762__;!!PvDODwlR4mBZyAb0!TJx5pTp_o-hhxhSprzQOv_6QSBNYmJ=
oh1Nzus_7wdo5w1UBXqIoOqY5ty-DiO-u2wXl1tuEpia221CneEgqz99Fta8BQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115049779=
/7559972__;!!PvDODwlR4mBZyAb0!TJx5pTp_o-hhxhSprzQOv_6QSBNYmJoh1Nzus_7wdo5w1=
UBXqIoOqY5ty-DiO-u2wXl1tuEpia221CneEgqz9wjmZWAA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Sep 03 2025 - 13:41:25 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 01 2025 - 14:23:05 MST