Re: [AMMRL] vacuum attachment for He transfer line

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 3 Sep 2025 23:48:41 +0200

Hi,

I might add that it would be nice to use a two-stage roughing pump which can
achieve on the order of 1-3 mbar whereas quite a few single-stage pumps
achieve only around 1 mbar which may stretch it a bit.

Yes to using a vacuum meter! 8-)

Btw. the pumping adapter on your helium transfer line looks strikingly like
the one used in some nitrogen dewars, so there may be a dual benefit from
getting this adapter ;-)

Cheers
Bert

> Am 03.09.2025 um 21:53 schrieb Fulton, Luke via groups.io:
>
> Hi Brian,
>
> Roughing pump should be fine, no need for a turbo.
> The biggest QOL improvement comes from attaching a vacuum gauge. Indispensable
> for knowing achievable end point, measuring vacuum progress, and re-testing
> for softening. Unfortunately they run upwards of $800. I'd check with coworkers
> before buying new; any synthetic lab doing schlenk work would have several in use.
>
> Kind regards,
> Luke
>
>
> Luke Fulton, PhD
> BLDG R003
> NMR Core Facility Director
> Unit 3060
> COR2E & Department of Chemistry
> 55 N Eagleville Road
> University of Connecticut
> Storrs, CT 06279
>
> email: fko24003_at_uconn.edu <mailto:fko24003_at_uconn.edu>
> (alias): luke.fulton_at_uconn.edu <mailto:luke.fulton_at_uconn.edu>
> mobile: (603) 953-5275
> Office: (860) 486-4069

> From: main_at_ammrl.groups.io on behalf of Brian Antalek via groups.io
> Sent: Wednesday, September 3, 2025 1:46 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] vacuum attachment for He transfer line
>
>
> I have a He transfer line that has softened and needs to be pumped down.
> It's from a 2005 ish Bruker AVII 300 MHz instrument and the vacuum connector
> piece cannot be found (see attached photo). Does anyone know where I can
> find one? Also, will a rough pump be enough, in general?
>
> Thanks
> Brian
>
> Brian Antalek
> Bausch + Lomb, Inc.
> 1400 N. Goodman Street
> Rochester, NY 14609
>
> brian.antalek_at_bausch.com <mailto:brian.antalek_at_bausch.com>
> (585) 752-8452 mobile
> (585) 666-1127 Teams
>
> <IMG_6465.jpeg>
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
MestreLab NMR Software (http://www.mestrelab.com)
Barthel HF Amplifiers (http://www.barthel-hf.de)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2763): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2763__;!!PvDODwlR4mBZyAb0!Tj82N89q7W3oJSDwK8_JyynVYG5KnX=
NZzkeC_fVjxEfbuONDe9UcbQRdiC7YPDPkuHg_q2azc4f2S1JWT948oA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115049779=
/7559972__;!!PvDODwlR4mBZyAb0!Tj82N89q7W3oJSDwK8_JyynVYG5KnXNZzkeC_fVjxEfbu=
ONDe9UcbQRdiC7YPDPkuHg_q2azc4f2S1LPEghfPw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Sep 03 2025 - 14:53:59 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 01 2025 - 14:23:05 MST