Hi Mike,
Yes, the helium is venting to atmosphere now. The normal boil off is
between 5-15 cc/hr depending on the day and temperature of the room.
Now it's at 15-17 cc/hr. The normal daily lost is 0.5-0.6 %. Now
it's at 0.78% daily. Normal weekly lost is 4-5%. Now is about 5-6%.
I think it has stabilized. I have the O-ring tape with electrical tape
and keeping record of the boil-off. I am crossing my fingers and hoping
it doesn't get worst, until we get a full assessment.
Best,
Hla
-=-=-=-=-=-=-=-=-=-=-=-