The easiest thing I would suggest is to run a quick 2H 1D and see what it
looks like. The Tol lock is expecting 4 peaks. If the shim is way off, maybe
it's having a hard time getting started?


[The Ohio State University]
Dr Alexandar L Hansen<https://osu.academia.edu/AlexandarHansen >
Lead Research Scientist
CCIC & Gateway NMR Facility
R224 CBEC, 151 W Woodruff Ave, Columbus, OH 43210
614.292.7133 Office / 716.341.8476 Mobile
hansen.434_at_osu.edu
ccic.ohio-state.edu


Pronouns: he/him/his

“Who's ready to make some science?” – Cave Johnson



________________________________
> From: main_at_ammrl.groups.io on behalf of Joseph Sachleben
> Sent: Friday, September 27, 2024 1:15 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] lock error on Bruker AVANCE III HD

HI All,

I have a weird lock problem on my 600 MHz Bruker AVANCE IIIHD system.
It won’t lock d8-toluene (the same sample locks fine on my 500 AVANCE III
system).
I get an error.

“Not all additional peaks could be found”.

It will lock D2O, CDCl3, d6-benzene, and even d8-toluene if I tell it it is
benzene. A screenshot of the error is below.

Thanks for your help.

Joe S.


Joseph R. Sachleben, Ph. D.
Technical Director Biomolecular NMR Facility
The Division of Biological Sciences
929 E. 57th St.
Chicago, IL 60637

(773)834-9866
jsachleben_at_uchicago.edu


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1657): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1657__;!!PvDODwlR4mBZyAb0!RP___E1Qhjox-xL1AAdwXiQDrOODqn=
8qgoresRkmlyOKrj61nx4GQOEJejuhDn5cN_d6W80GY62biiByuYM$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108691017=
/7559972__;!!PvDODwlR4mBZyAb0!RP___E1Qhjox-xL1AAdwXiQDrOODqn8qgoresRkmlyOKr=
j61nx4GQOEJejuhDn5cN_d6W80GY62bidRboyg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Sep 27 2024 - 11:41:44 MST

This archive was generated by hypermail 2.4.0 : Tue Oct 01 2024 - 14:06:56 MST