Hi,

primary issue is, not the error message.
Primary issue is an apparent weak lock power.

Step by step. It's an Avance III console and a QNP cryoprobe. Running with TopSpin 3.6.2
In principle everything is fine, exept the need of an an exceptional
High lock power to get a reasonable lock signal.

After some investigations there seems to be a good chance, that the lock
preamplifier inside the cryo probe and/or the T/R switch is broken. It's
possible to bypass the built in preamplifier and use the external HPPR
preamplifier.
This would decrease the S/N, of course.

I made this successfully in the past, about 10...15 years ago. The necessary
tool is UniTool and here the bsms Option. Using the "set Force-State" option
it should be possible to activate/deactivate the built in preamplifier.
Every of the displayed UniTool options works well, except "G" and "S". If you
press one of the letters, you get the
Error message
     Illegal '%' escape in menu order string
Tried lower and capital letters.

I strongly assume, it's a tiny UniTool bug only. Maybe UniTool of another
TopSpin versions doesn't show this error?

A funny challenge for an retired old man.

Greetings

Rainer

PS: If you don't know, maybe this webpage is of interest.
https://www.nmr.tips/indexe.html



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1656): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1656__;!!PvDODwlR4mBZyAb0!XuTD6C5NXAzdLgxgT5gzsY0xQjBi_B=
CpODQz1wKUMmFlCCf_zB-ZMqvjcNY9YfegSkmp7_5mozAE_Q$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108691532=
/7559972__;!!PvDODwlR4mBZyAb0!XuTD6C5NXAzdLgxgT5gzsY0xQjBi_BCpODQz1wKUMmFlC=
Cf_zB-ZMqvjcNY9YfegSkmp7_6gS74IUQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Sep 27 2024 - 11:06:41 MST

This archive was generated by hypermail 2.4.0 : Tue Oct 01 2024 - 14:06:56 MST