Re: [AMMRL] BACs downwards motion failed

From: Tim Claridge <tclaridge_at_exscientia.co.uk>
Date: Thu, 26 Sep 2024 12:47:22 +0000

As described in Ulrich’s mail- this problem is (nearly) always associated
with dirt on the contacts within the rectangular piston block- cleaning the
piston arm itself is unlikely to resolve the issue. If the 4 allen bolts
are remove from the end of the block it is possible to access and clean the
internal contacts as described. These touch on a (blue) rubberised ring that
is conductive (impregnated) and when compressed (by the upward or downward
motion of the piston) will complete the contact that indicates the motion
is complete. Dirt on the contacts and/or the rubberised ring can lead to the
upward/downward motion failed error. We used to keep a spool of the blue rubber
tubing so we could replace this periodically to resolve these issues, in
addition to cleaning. Too much grease on the piston can also cause grease to
accumulate on the rubber which can exacerbate the problem, so go easy on
the piston grease.

Good luck.

Tim

Tim Claridge
Associate Director, NMR Analytics
+44 (0) 7842 000508
Exscientia

155 Brook Drive, Milton Park, Abingdon, OX14 4SD

Prix Galien International, Best Digital Health Solution 2024
UK Queen’s Award for Enterprise in Innovation 2021



> From: main_at_ammrl.groups.io on behalf of Craig Grimmer via groups.io
> Date: Thursday, 26 September 2024 at 12:21
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] BACs downwards motion failed

The little pots of grease are labelled "Horiz-Zyl" and "Vertikal-Zyl


--
*Please note that this email communication and any attachments to it may
contain confidential or privileged information and so circulation should be
restricted to the addressee’s organisation. If you are not the addressee,
please let us know by return email, then delete all copies of this email.
**
Personal data that you share with us will be stored and used in
accordance with our Privacy Policy which can be accessed on our website.*
*
Exscientia plc is a company incorporated in England and Wales with
registered number 13483814 and its registered office at The Schrodinger
Building, Oxford Science Park, Oxford, Oxfordshire, United Kingdom, OX44GE*



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1647): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1647__;!!PvDODwlR4mBZyAb0!Q3-oe21asFrlA_OjYd-E5ESlwSPYLw=
XsNNTJnM00pZ5ovkOh5gXYOBQey1jZyLTP94WTUMA7RkAGdCU0eg2toJY$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108655772=
/7559972__;!!PvDODwlR4mBZyAb0!Q3-oe21asFrlA_OjYd-E5ESlwSPYLwXsNNTJnM00pZ5ov=
kOh5gXYOBQey1jZyLTP94WTUMA7RkAGdCU04PkCu80$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Sep 26 2024 - 05:47:37 MST

This archive was generated by hypermail 2.4.0 : Tue Oct 01 2024 - 14:06:56 MST