As Ken says the companies offer little help with this. Each site is very
different so without really knowing your setup this is hard to answer.
I have talked to a couple of sites that have these all metal systems (no bag).
One issue is that they connected everything up and then put the 2in line into
a 1/4in hole on the tank. So when attempting to fill the system they had lots
of back pressure. So assuming your systems are plumbed correctly I think the
PSI value that your compressor turns on at 2.5 needs to be lowered. Magnets
will not vent normally at that PSI. My understanding is that when there is a
small pressure in the lines and tank the compressor should cycle on and off
(can potentially pull on your magnets this is why back pressure controllers
and such are used). During a fill the lines can not become too pressurized
or else normal boil-off will not happen, this would be an issue for the
second magnet not being filled.
Hopefully you will get more info from those who have this type of system. We
have high pressure here with a gas bag.
Greg
Gregory P. Wylie, Ph.D
NMR Facility Manager
Texas A&M University
Department of Chemistry
P.O. Box 30012
College Station, TX 77843-3255
gpwylie_at_tamu.edu
979.458.0705 (voice)
979.845.4719 (fax)
706.206.0007 (cell)
http://nmr.tamu.edu
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#545): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/545__;!!PvDODwlR4mBZyAb0!XwY9J4Ik-q3JGXu8thBilnHox7YwI6dx=
qyo5yrprX6VkxezKHDD6oRzvOCzkla8OR4ZWjJDTfiwm1zc-UhMuBPclFIae$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/101199948=
/7559972__;!!PvDODwlR4mBZyAb0!XwY9J4Ik-q3JGXu8thBilnHox7YwI6dxqyo5yrprX6Vkx=
ezKHDD6oRzvOCzkla8OR4ZWjJDTfiwm1zc-UhMuBO266V7R$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
From owner-ammrl_at_ammrl.groups.io Wed, 6 Sep 2023 13:52:49
Received: by 2002:a05:7022:1085:b0:6a:a843:e96f with SMTP id ep5csp148264dlb;
Wed, 6 Sep 2023 13:52:49 -0700 (PDT)
X-Google-Smtp-Source: AGHT+IElOiVTMd6ixIBBDf4ZW9gg+tX0sNKA5rFmZBAEprsxvjoLHupoNYZsskEebJgLTU0Amy0i
X-Received: by 2002:a25:cbce:0:b0:d74:cdb:adb8 with SMTP id b197-20020a25cbce000000b00d740cdbadb8mr16352502ybg.28.1694033569392;
Wed, 06 Sep 2023 13:52:49 -0700 (PDT)
ARC-Seal: i=1; a=rsa-sha256; t=1694033569; cv=none;
d=google.com; s=arc-20160816;
b=BoRzhHXEWbhgXUoZac4qdFTGIqtHC2FCn5YqamOHavikEDT85lfSxrh4mGAJWr/SnD
bRpOfXhHrpLvRycsWNa+8vBxBvMxsC9olf9UYEC9rOgV3chB6WpJjucnwSi988N/Qsgb
hhLXK09PMDmJUyho9F7AHxwR1lX5nhB9bI3QTK2PBvoof1ICLH1DQvGnfJnGGilBYPOI
2p8NeExS2IdB6R8UO1uCP7yvf04LKC/B7G8rGzwsUQo1QoQKq1PL27nvNsalCvw5lxsv
J4AqACf4CH8H5p3OFlcT4MO952b1VIym/njBVvBwY3o4W1aEV3xdmCP1BOuGDdQ21/p2
OjIA==
ARC-Message-Signature: i=1; a=rsa-sha256; c=relaxed/relaxed; d=google.com; s=arc-20160816;
h=mime-version:list-unsubscribe:list-unsubscribe-post:reply-to
:delivered-to:mailing-list:list-id:sender:list-help:list-subscribe
:precedence:to:subject:message-id:date:from:in-reply-to:references
:dkim-signature;
bh=yMRq3KEJG1K0krjhtN2C5paUP50UjjrEJWg5F+XnhXo=;
fh=xC55mW4WXjmdiwtkPJUr4APdxMPTmW+ubeamlxP9uX0=;
b=Ll3MkV8ao2FKdi93AN47xTXNaQhaswKkLpbzMxkIYeFNdC7sDQG5EOPq8vzhmzB64S
pFUoLZli9VYp0cAw6bY5yfuaxeFKPY2FlyiZk229NQMKMEXqL6/+yuM8fc77ArsC03Bc
iVu9BnQoucpb/J1tVfywODrW+TZgZniLQBsZWrS7sz8XPvAI49b54zG8Va+h/wx16Bgl
NJkwEXVcNvsR398zTX3hWG3S+4zOAR4rKDHlXMUMa9PfqLf7gYQ06OzPGaokpnpPcSNT
lSN3yANN0mjnUGaMEAIPUy8D6/7eMpC+lsPPNxp3MBxauS+dZL+7fOo68DfEXCWTB1R7
0DzA==
ARC-Authentication-Results: i=1; mx.google.com;
dkim=neutral (body hash did not verify) header.i=_at_groups.io header.s=20140610 header.b=ohAUoR9K;
spf=pass (google.com: domain of bounce+115296+546+7559972+12177410_at_groups.io designates 66.175.222.108 as permitted sender) smtp.mailfrom=bounce+115296+546+7559972+12177410_at_groups.io
Return-Path: <bounce+115296+546+7559972+12177410_at_groups.io>
Received: from mx0b-007a6701.pphosted.com (mx0b-007a6701.pphosted.com. [205.220.181.169])
by mx.google.com with ESMTPS id y8-20020a25ad08000000b00d7e3a0f32f8si14577916ybi.240.2023.09.06.13.52.49
for <walt_at_hawaii.edu>
(version=TLS1_2 cipher=ECDHE-RSA-AES128-GCM-SHA256 bits=128/128);
Wed, 06 Sep 2023 13:52:49 -0700 (PDT)
Received-SPF: pass (google.com: domain of bounce+115296+546+7559972+12177410_at_groups.io designates 66.175.222.108 as permitted sender) client-ip=66.175.222.108;
Authentication-Results: mx.google.com;
dkim=neutral (body hash did not verify) header.i=_at_groups.io header.s=20140610 header.b=ohAUoR9K;
spf=pass (google.com: domain of bounce+115296+546+7559972+12177410_at_groups.io designates 66.175.222.108 as permitted sender) smtp.mailfrom=bounce+115296+546+7559972+12177410_at_groups.io
Received: from pps.filterd (m0313027.ppops.net [127.0.0.1])
by mx0b-007a6701.pphosted.com (8.17.1.19/8.17.1.19) with ESMTP id 386B4COW025098
for <walt_at_hawaii.edu>; Wed, 6 Sep 2023 10:52:49 -1000
Authentication-Results: ppops.net;
spf=pass smtp.mailfrom=bounce+115296+546+7559972+12177410_at_groups.io;
dkim=pass header.s=20140610 header.d=groups.io;
dmarc=none
Received: from mail02.groups.io (mail02.groups.io [66.175.222.108])
by mx0b-007a6701.pphosted.com (PPS) with ESMTPS id 3sxg2dw7hv-1
(version=TLSv1.2 cipher=ECDHE-RSA-AES128-GCM-SHA256 bits=128 verify=NOT)
for <walt_at_hawaii.edu>; Wed, 06 Sep 2023 10:52:48 -1000
DKIM-Signature: a=rsa-sha256; bh=M4pSJSccXN1XvIvQrgq+JdpntktUW7PQdFfOI4GgfpY=;
c=relaxed/simple; d=groups.io;
h=MIME-Version:References:In-Reply-To:From:Date:Message-ID:Subject:To:Precedence:List-Subscribe:List-Help:Sender:List-Id:Mailing-List:Delivered-To:Reply-To:List-Unsubscribe-Post:List-Unsubscribe:Content-Type;
s=20140610; t=1694033567; v=1;
b=ohAUoR9Kvsc+bYAQLQChLzFCy+CqMR640zFhYRNgIMFy5cc3Gj6OtoB47jogOBFmuAju9s4Z
btd14N4QHOeU4Zu0kXsVNxvgPZZJ2CkDR2w7yloK5jj1mWaZNAux29ILZBO5pQ7u8HmOrxqVqwk
gwFXA/SOSeiPztsfbSh0Ff6w=
X-Received: by 127.0.0.2 with SMTP id UGOjYY7561307x9qruvuvJP2; Wed, 06 Sep 2023 13:52:47 -0700
X-Received: from mail-yb1-f169.google.com (mail-yb1-f169.google.com [209.85.219.169])
by mx.groups.io with SMTP id smtpd.web10.1190.1694033566321130056
for <main_at_ammrl.groups.io>;
Wed, 06 Sep 2023 13:52:46 -0700
X-Received: by mail-yb1-f169.google.com with SMTP id 3f1490d57ef6-d7e9d849bdfso240749276.3
for <main_at_ammrl.groups.io>; Wed, 06 Sep 2023 13:52:46 -0700 (PDT)
X-Gm-Message-State: 4DkeRzoJqUaruAmtgl6QWELMx7559972AA=
X-Received: by 2002:a25:40a:0:b0:d7f:7690:d865 with SMTP id
10-20020a25040a000000b00d7f7690d865mr5721425ybe.6.1694033565266; Wed, 06 Sep
2023 13:52:45 -0700 (PDT)
References: <3D8FA702-BCC4-4265-98C7-2AB37DA9FA90_at_mines.edu> <1064.1694032678769000831_at_groups.io>
In-Reply-To: <1064.1694032678769000831_at_groups.io>
From: "Xiaoyan Sun via groups.io" <xsun9=ncsu.edu_at_groups.io>
Date: Wed, 6 Sep 2023 16:52:34 -0400
Message-ID: <CAOKgmAJ2gtpDY=SAFvgcJN6qLoWWQa--3M6bfZ8VW=D93tHj=w_at_mail.gmail.com>
Subject: Re: [AMMRL] CryoMech Compact Helium Recovery System
To: main_at_ammrl.groups.io
Precedence: Bulk
List-Subscribe: <mailto:main+subscribe_at_ammrl.groups.io>
List-Help: <mailto:main+help_at_ammrl.groups.io>
Sender: main_at_ammrl.groups.io
List-Id: <main.ammrl.groups.io>
Mailing-List: list main_at_ammrl.groups.io; contact main+owner_at_ammrl.groups.io
Delivered-To: mailing list main_at_ammrl.groups.io
Reply-To: main_at_ammrl.groups.io
List-Unsubscribe-Post: List-Unsubscribe=One-Click
List-Unsubscribe: <https://ammrl.groups.io/g/main/leave/12177410/7559972/1125599165/plugh>
Content-Type: multipart/alternative; boundary="000000000000e5663e0604b6ec21"
X-Proofpoint-GUID: 6VXzsnIo2ntoabxh_TUU7Mlfk8hzex-w
X-Proofpoint-ORIG-GUID: 6VXzsnIo2ntoabxh_TUU7Mlfk8hzex-w
X-CLX-Shades: MLX
X-CLX-Response: 1TFkXGB4bEQpMehcYGB8RCllEF2BFbkNLR0NdaxxMEQpYWBdubl1TGmdCbmR 7bhEKeE4XaGxnQRJLaXIYfXMRCnhLF25uXVMaZ0JuZHtuEQp5TBdmHmlNYx59ZlkFTxEKQ0gXBx 4YGBEKQ1kXBxsfGhEKQ0kXGgQaGhoRCllNF2dmchEKWUkXGnEaEBp3BhgYG3EbHBsQGncGGBoGG
hEKWV4XbGxmEQpJRhdfRENcT1hZQ15TQktdS0NDdUJFWV5PThEKSUcXeE9NEQpDThccfHJQWURj RRhEXkVLSFJCdX5/fx1nRkxBEkJQT1IHXREKWFwXHwQaBBkYHQUbGgQbGxoEGxkeBBkZEBseGh8 aEQpeWRdJZkVZHREKTVwXGRwcEQpMWhdobVJNUxEKTU4XaBEKTEYXb2trXmtrEQpCTxdtHlJHAR
hwSGZyaxEKQ1oXHBwEGx0fBBgYGAQbGhIRCkJeFxsRCkJGF2BFbkNLR0NdaxxMEQpCRxdiXV4ac EdzTmV6QREKQlwXGxEKXk4XGxEKQksXZRtuG1xGex1jSGcRCkJJF2UbbhtcRnsdY0hnEQpCRRd6 a0BdG2loekReAREKQk4XZRtuG1xGex1jSGcRCkJMF3pZTklwT1McemxdEQpCbBdmZmhHWW5lU0d
zYBEKQkAXentrUEAeYkdfcE0RCkJYF25waGNYe258G1NMEQpaWBccEQp5QxdvQh1EQX8YQ1haXx EKWUsXHhwSHxEKWksXHhwSHxEKcGcXYm9uXEceWE5lYGIQGhEKcGgXYxh6W0R7AW0FbkQQGhEKc GgXbRp6eltDUkZHaHoQGhEKcGgXbElvXl1TaH5zEmcQGhEKcGgXZHoca15zZHwfXV8QGhEKcGgX
b3MfYWhyU3paGWEQGhEKcGgXYkBibhlrXGhyYXwQGhEKcGgXbGUbGgEfa2FIb2QQGhEKcGgXaUh zQEkTU1trWBgQGhEKcGcXemt9HUtyUnl/b0AQGhEKcH0XaHAFb0ROb2BkclIQGhEKcGcXbB1eS0 5gbx1yE0AQGhEKcGcXbXtdc0J/Un5NQGYQGhEKcH0XYURjeUEZZWFZf0MQGhEKcH0XZh15XWFdR
mhfYXsQGhEKcH0XZEFkGgFkYXlMH2IQGhEKcH0XZm9PYV1SWm1LE08QGhEKcGcXZWZZbF5HaB9P X30QHhIRCnB9F2dsH1B/Q05GU0MBEBoRCnBnF2xsXUFOHU4dXk9lEBoRCnB9F29dTm4ZQ05OU1p /EBoRCnBnF2YBRx5PHWJTYx9JEB4SEQpwfRdseHweW1IaZV4cYhAeEhEKcH0XYW5IAVpzfG4ZYV
sQGhEKcH8XYkZTeGZoaR1hTEsQGxkaEQpwXxdocAVvRE5vYGRyUhAaEQpwfxdsHHpaWklBf0xGb BASGBEKcF8XZXxbTmJpUk9/ek4QGhEKcF8XZxhMSxpsG0cYBXMQGhEKcH8XbBlZAW1kEmZFGX4Q GxsZEQpwXxdmWltcb0dLbnJOUxAeEhEKcF8XbFx7Em1vE09wH0UQGxscEQpwXxdofE9gUG1oa1k
eYxAeEhEKcF8XbHh8HltSGmVeHGIQHhIRCnBfF2sFGmh9XVBSYHATEB4SEQpwbBdtYnlYf0gZHU BFbBAaEQptfhcaEQpYTRdLESA=
MIME-Version: 1.0
X-Proofpoint-Virus-Version: vendor=baseguard
engine=ICAP:2.0.267,Aquarius:18.0.957,Hydra:6.0.601,FMLib:17.11.176.26
definitions=2023-09-06_06,2023-09-05_01,2023-05-22_02
X-Proofpoint-Spam-Details: rule=inbound_notspam policy=inbound score=0 bulkscore=0 lowpriorityscore=0
suspectscore=0 spamscore=0 mlxlogscore=999 priorityscore=225 phishscore=0
malwarescore=0 mlxscore=0 adultscore=0 unknownsenderscore=20
impostorscore=0 clxscore=241 classifier=spam adjust=0 reason=mlx
scancount=1 engine=8.12.0-2308100000 definitions=main-2309060181
domainage_hfrom=4685 domainage_replyto=4685
--000000000000e5663e0604b6ec21
Content-Type: text/plain; charset="UTF-8"
Content-Transfer-Encoding: quoted-printable
Any recommendation on the brand and type of back pressure regulator?
On Wed, Sep 6, 2023 at 4:38=E2=80=AFPM <gregory.wylie_at_chem.tamu.edu> wrote:
> HI Yuan
>
> As Ken says the companies offer little help with this. Each site is very
> different so without really knowing your setup this is hard to answer.
>
> I have talked to a couple of sites that have these all metal systems (no
> bag). One issue is that they connected everything up and then put the 2in
> line into a 1/4in hole on the tank. So when attempting to fill the system
> they had lots of back pressure. So assuming your systems are plumbed
> correctly I think the PSI value that your compressor turns on at 2.5 needs
> to be lowered. Magnets will not vent normally at that PSI. My understanding
> is that when there is a small pressure in the lines and tank the compressor
> should cycle on and off (can potentially pull on your magnets this is why
> back pressure controllers and such are used). During a fill the lines can
> not become too pressurized or else normal boil-off will not happen, this
> would be an issue for the second magnet not being filled.
>
> Hopefully you will get more info from those who have this type of system.
> We have high pressure here with a gas bag.
>
> Greg
> Gregory P. Wylie, Ph.D
> NMR Facility Manager
> Texas A&M University
> Department of Chemistry
> P.O. Box 30012
> College Station, TX 77843-3255
> gpwylie_at_tamu.edu
> 979.458.0705 (voice)
> 979.845.4719 (fax)
> 706.206.0007 (cell)
> http://nmr.tamu.edu
>
>
>
--=20
*Jennifer Sun*
Senior Research Scholar , Molecular Education, Technology and Research Innovation Center
Joint Research Assistant Professor, Department of Chemistry
North Carolina State University
2620 Yarbrough Drive
Raleigh, NC 27607
(919) 515-1160
*xsun9_at_ncsu.edu <xsun9_at_ncsu.edu>*
https://research.ncsu.edu/metric/
https://labs.ncsu.edu/
Please citing METRIC to support us:
All NMR measurements were made in the Molecular Education, Technology and
Research Innovation Center (METRIC) at NC State University.
-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#546): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/546__;!!PvDODwlR4mBZyAb0!QdQdGZAFSbPPNU-qBKUqO9eRyzuUMxLM=
FcKqItUCrnv_rvZY4sIHANFJJoyVm8a5aDX7idblscoHwBR93MvImjUSUQ$=20
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/101199948=
/7559972__;!!PvDODwlR4mBZyAb0!QdQdGZAFSbPPNU-qBKUqO9eRyzuUMxLMFcKqItUCrnv_r=
vZY4sIHANFJJoyVm8a5aDX7idblscoHwBR93MtOHGSOwQ$=20
Group Owner: main+owner_at_ammrl.groups.io
-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-=3D-
--000000000000e5663e0604b6ec21
Content-Type: text/html; charset="utf-8"
Content-Transfer-Encoding: quoted-printable
<div dir=3D"ltr">Any recommendation=C2=A0on the brand and type of back pres=
sure regulator?</div><br/><div class=3D"gmail_quote"><div dir=3D"ltr" class=
=3D"gmail_attr">On Wed, Sep 6, 2023 at 4:38=E2=80=AFPM <<a href=3D"mailt=
o:gregory.wylie_at_chem.tamu.edu">gregory.wylie_at_chem.tamu.edu</a>> wrote:<b=
r/></div><blockquote class=3D"gmail_quote" style=3D"margin:0px 0px 0px 0.8e=
x; border-left:1px solid rgb(204,204,204); padding-left:1ex;">HI Yuan<br/><=
br/>As Ken says the companies offer little help with this. Each site is ver=
y different so without really knowing your setup this is hard to answer.<br=
/><br/>I have talked to a couple of sites that have these all metal systems=
(no bag). One issue is that they connected everything up and then put the =
2in line into a 1/4in hole on the tank. So when attempting to fill the syst=
em they had lots of back pressure. So assuming your systems are plumbed cor=
rectly I think the PSI value that your compressor turns on at 2.5 needs to =
be lowered. Magnets will not vent normally at that PSI. My understanding is=
that when there is a small pressure in the lines and tank the compressor s=
hould cycle on and off (can potentially pull on your magnets this is why ba=
ck pressure controllers and such are used). During a fill the lines can not=
become too pressurized or else normal boil-off will not happen, this would=
be an issue for the second magnet not being filled.<br/><br/>Hopefully you=
will get more info from those who have this type of system. We have high p=
ressure here with a gas bag.<br/><br/>Greg<br/>
<div>Gregory P. Wylie, Ph.D</div>
<div>NMR Facility Manager<br/>Texas A&M University<br/>Department of Ch=
emistry<br/>P.O. Box 30012<br/>College Station, TX =C2=A077843-3255<br/><a =
href=3D"mailto:gpwylie_at_tamu.edu" rel=3D"noopener" target=3D"_blank">gpwylie=
_at_tamu.edu</a><br/>979.458.0705 (voice)<br/>979.845.4719 (fax)<br/>706.206.0=
007 (cell)<br/><a href=3D"
https://urldefense.com/v3/__http://nmr.tamu.edu/_=
_;!!PvDODwlR4mBZyAb0!QdQdGZAFSbPPNU-qBKUqO9eRyzuUMxLMFcKqItUCrnv_rvZY4sIHAN=
FJJoyVm8a5aDX7idblscoHwBR93MvoWg497w$" rel=3D"noopener" target=3D"_blank">h=
ttp://nmr.tamu.edu</a></div>
<p></p><p></p></blockquote></div><br clear=3D"all"/><div><br/></div><span c=
lass=3D"gmail_signature_prefix">-- </span><br/><div dir=3D"ltr" class=3D"gm=
ail_signature"><div dir=3D"ltr"><div><div dir=3D"ltr"><div dir=3D"ltr"><p c=
lass=3D"MsoNormal" style=3D"margin:0in 0in 8pt; line-height:11.75pt; border=
:none; padding:0in;"><defangedfont face=3D"georgia, serif"><span style=3D"b=
ackground:rgb(255,255,255);"><b>Jennifer Sun</b></span><i><u></u></i></defa=
ngedfont></p><p class=3D"MsoNormal" style=3D"margin:0in 0in 8pt; line-heigh=
t:11.75pt; border:none; padding:0in;"><defangedfont face=3D"georgia, serif"=
><span style=3D""><defangedfont size=3D"2">Senior Research Scholar=C2=A0</d=
efangedfont></span><span style=3D"">, Molecular Education, Technology and R=
esearch Innovation Center</span></defangedfont></p><p class=3D"MsoNormal" s=
tyle=3D"margin:0in 0in 8pt; line-height:11.75pt; border:none; padding:0in;"=
><defangedfont face=3D"georgia, serif"><span style=3D"">Joint Research Assi=
stant Professor</span><span style=3D"">, Department of Chemistry</span><u><=
/u></defangedfont></p><p class=3D"MsoNormal" style=3D"margin:0in 0in 8pt; l=
ine-height:11.75pt; border:none; padding:0in;"><span style=3D""><defangedfo=
nt face=3D"georgia, serif">North Carolina State University</defangedfont></=
span></p><p class=3D"MsoNormal" style=3D"margin:0in 0in 8pt; line-height:11=
.75pt; border:none; padding:0in;"><span style=3D""><defangedfont face=3D"ge=
orgia, serif">2620 Yarbrough Drive</defangedfont></span></p><p class=3D"Mso=
Normal" style=3D"margin:0in 0in 8pt; line-height:11.75pt; border:none; padd=
ing:0in;"><span style=3D""><defangedfont face=3D"georgia, serif">Raleigh, N=
C 27607</defangedfont></span></p><p class=3D"MsoNormal" style=3D"margin:0in=
0in 8pt; line-height:11.75pt; border:none; padding:0in;"><span style=3D"">=
<defangedfont face=3D"georgia, serif">(919) 515-1160</defangedfont></span><=
/p><p class=3D"MsoNormal" style=3D"margin:0in 0in 8pt; line-height:11.75pt;=
border:none; padding:0in;"><defangedfont color=3D"#666666"><i><u><a href=
=3D"mailto:xsun9_at_ncsu.edu" style=3D"" target=3D"_blank">xsun9_at_ncsu.edu</a><=
/u></i></defangedfont></p><p class=3D"MsoNormal" style=3D"margin:0in 0in 8p=
t; line-height:11.75pt; border:none; padding:0in;"><a href=3D"
https://urlde=
fense.com/v3/__
https://research.ncsu.edu/metric/__;!!PvDODwlR4mBZyAb0!QdQdG=
ZAFSbPPNU-qBKUqO9eRyzuUMxLMFcKqItUCrnv_rvZY4sIHANFJJoyVm8a5aDX7idblscoHwBR9=
3MtZxtqoUQ$" style=3D"" target=3D"_blank"><span style=3D"">
https://research=
.ncsu.edu/metric/</span></a></p><div dir=3D"ltr"><span style=3D""><a href=
=3D"
https://urldefense.com/v3/__https://labs.ncsu.edu/__;!!PvDODwlR4mBZyAb0=
!QdQdGZAFSbPPNU-qBKUqO9eRyzuUMxLMFcKqItUCrnv_rvZY4sIHANFJJoyVm8a5aDX7idblsc=
oHwBR93MvzbfhUzg$" style=3D"" target=3D"_blank">
https://labs.ncsu.edu/</a><=
/span></div><div dir=3D"ltr"><br/></div><div dir=3D"ltr"><span style=3D"">P=
lease citing METRIC to support us:</span><div><span style=3D"">All=C2=A0</s=
pan><span style=3D"">NMR</span><span style=3D"">=C2=A0measurements were mad=
e in the Molecular Education, Technology and Research Innovation Center (ME=
TRIC) at NC State University.<br/></span><br/><br/></div></div></div></div>=
</div></div></div>
<div width=3D"1" style=3D"color:white;clear:both">_._,_._,_</div>
<hr>
Groups.io Links:<p>
=20=20
You receive all messages sent to this group.
=20=20
=20=20
<p>
<a target=3D"_blank" href=3D"
https://urldefense.com/v3/__https://ammrl.grou=
ps.io/g/main/message/546__;!!PvDODwlR4mBZyAb0!QdQdGZAFSbPPNU-qBKUqO9eRyzuUM=
xLMFcKqItUCrnv_rvZY4sIHANFJJoyVm8a5aDX7idblscoHwBR93MvImjUSUQ$">View/Reply =
Online (#546)</a> |
<a target=3D"_blank" href=3D"mailto:main_at_ammrl.groups.io?subject=3DRe:%20=
Re%3A%20%5BAMMRL%5D%20CryoMech%20Compact%20Helium%20Recovery%20System">Repl=
y To Group</a>
=20=20
| <a target=3D"_blank" href=3D"mailto:xsun9_at_ncsu.edu?subject=3DPrivate:=
%20Re:%20Re%3A%20%5BAMMRL%5D%20CryoMech%20Compact%20Helium%20Recovery%20Sys=
tem">Reply To Sender</a>
=20=20
|
<a target=3D"_blank" href=3D"
https://urldefense.com/v3/__https://groups.i=
o/mt/101199948/7559972__;!!PvDODwlR4mBZyAb0!QdQdGZAFSbPPNU-qBKUqO9eRyzuUMxL=
MFcKqItUCrnv_rvZY4sIHANFJJoyVm8a5aDX7idblscoHwBR93MtOHGSOwQ$">Mute This Top=
ic</a>
| <a href=3D"
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/pos=
t__;!!PvDODwlR4mBZyAb0!QdQdGZAFSbPPNU-qBKUqO9eRyzuUMxLMFcKqItUCrnv_rvZY4sIH=
ANFJJoyVm8a5aDX7idblscoHwBR93Mu1V_LMpg$">New Topic</a>
<br>
<a href=3D"
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/edits=
ub/7559972__;!!PvDODwlR4mBZyAb0!QdQdGZAFSbPPNU-qBKUqO9eRyzuUMxLMFcKqItUCrnv=
_rvZY4sIHANFJJoyVm8a5aDX7idblscoHwBR93Muw1nn7tg$">Your Subscription</a> |
<a href=3D"mailto:main+owner_at_ammrl.groups.io">Contact Group Owner</a> |
<div width=3D"1" style=3D"color:white;clear:both">_._,_._,_</div>
--000000000000e5663e0604b6ec21--
Received on Wed Sep 06 2023 - 13:38:01 MST