Hi Pedro,
Cisplatin was for liquid state.
2.5mm ss probe is on a different spectrometer
-Ravi
From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Pedro Aguiarvia groups.io
Sent: Wednesday, September 6, 2023 3:25 PM
To: main_at_ammrl.groups.io
Subject: Re: [AMMRL] 2.5mm spin problems
Hello Ravi,
I saw that you also asked a question about cis-platin. Are the samples you're
trying to spin in your 2.5mm MAS platinum containing compounds? If the density
of the compounds is very high you may have greater difficulty spinning.
You might want to try the drive cap/ends which have a longer portion inside
of the rotor to constrict the sample to centre-most part of the rotor.
Cheers,
Pedro
****************************************************************
Dr. Pedro M. Aguiar, MRSC (UK)
Coordonnateur - Laboratoire de RMN
Département de Chimie
Pavillon Roger-Gaudry, E-715
L'Université de Montréal
Montréal, QC H3C 3J7
Canada
Email: pedro.aguiar_at_umontreal.ca<mailto:pedro.aguiar_at_umontreal.ca>
Tel: 514-343-6111 ext. 31006
http://chimie.umontreal.ca/en/accueil/
https://sites.google.com/site/pedroaguiarresearch/
*****************************************************************
Je comprends que vos heures de travail ne soient peut-être pas identiques mes heures de travail.
S'il vous plaît répondre pendant vos heures de travail.
I appreciate that your working hours may not be the same as my working hours.
Please reply during your working hours.
On Sep 6, 2023, at 13:02, Garimella, Ravi <rgarimel_at_odu.edu<mailto:rgarimel_at_odu.edu>> wrote:
Hi spins,
Can you recommend any troubleshooting tips for spin problems ( not spinning
with good marking present at the bottom) for a 2.5 mm ssnmr probe?
Thank you,
_Best,
(Ravindranath Garimella PhD)
NMR Manager
College of Sciences Major Instrumentation Cluster (COSMIC) Facility
1104 Oceanography and Physical SCIENCES Bidg.,
4600 Elkhorn Ave.
Norfolk, VA 23529.
Email : rgarimel_at_odu.edu<mailto:rgarimel_at_odu.edu>
Ph : (757)683-3434
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#544): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/544__;!!PvDODwlR4mBZyAb0!RKHF5E3BOBUyP9gNiA8VNnw2n-k2Jd6B=
NqphtyTbPNwmG2xAxomZ6UWisA8qgZ-xYPpCgRAwCEWuj5aq$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/101198059=
/7559972__;!!PvDODwlR4mBZyAb0!RKHF5E3BOBUyP9gNiA8VNnw2n-k2Jd6BNqphtyTbPNwmG=
2xAxomZ6UWisA8qgZ-xYPpCgRAwCMiTwdA1$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Sep 06 2023 - 13:36:22 MST