Re: AMMRL: UPS IBFF problem fixed

From: Anbanandam, Asokan <asokan_at_ku.edu>
Date: Tue, 15 Sep 2015 17:46:15 +0000

Dear All:

Thank you so much for your quick reply to my request. I want to personally Thank the following members for suggestions and solutions to the UPS problem.


Martha Mortan (UNL, Lincoln, NE)

Jerry Hirschinger (Purdue univ., IN)

Jerry Hu (UCSB, CA)

Youlon Xia (UMN, MN): Youlin has excellent step by step protocol to fix UPS battery or fuse problem.

Thomas Eubanks (UTPA, PA??)

Wei Weicoff (MU, MO)

Valerie Robertson (Uof Guelph, CA)

Ben Ramirez (UIC, IL)

Bill Bartch 9BRUKER)


How did I solve the problem:

I simply followed Youlin's step by step protocol. After removing the screws at the front and back of the UPS, then removed the cover. I noticed that one of the fan was not spinning on normal speed as the other one. I simply used my finger to speed up the slow spinning fan, it came back to normal speed. This process simply cleared the error message on the UPS as well as cleared Low UPS message on the BMPC panel. Everything is working normal as of now. But, down the road, I may have to replace the slow spinning fan sooner than later.


Once again Thank You all for your help.

Asokan

?

***********************************
Dr. A. Asokan, Ph.D.
Director, Bio-NMR (CORE FACILITY)
Del Shankel Structural Biology Center
2034 Becker Dr. #1011
The University of Kansas
Lawrence, KS 66047
785-864-3746 (O)
785-864-8141 (F)
http://psf.cobre.ku.edu/BioNMRLab.htm
***********************************


________________________________
From: Anbanandam, Asokan
Sent: Monday, September 14, 2015 10:56 AM
To: ammrl_at_ammrl.org
Subject: AMMRL: UPS INTERNAL BATTERY FUSE FAILURE


Greetings All:

Our 800 MHz NMR backup power supply unit, GE digital LAN PRO 8-11 model is flashing warning LED on the front panel, prompting me to press the online alarm button to

see the warning message. It reads as "Internal battery fuse failure". Though it is a GE product, it is manufactured in Switzerland. According to the manual it has 20 battery pack. Has anyone had a similar problem and if so, how did you resolve it. How much it would cost to replace the battery pack. Any advise or suggestion is greatly appreciated.

Thanks

Asokan


***********************************
Dr. A. Asokan, Ph.D.
Director, Bio-NMR (CORE FACILITY)
Del Shankel Structural Biology Center
2034 Becker Dr. #1011
The University of Kansas
Lawrence, KS 66047
785-864-3746 (O)
785-864-8141 (F)
http://psf.cobre.ku.edu/BioNMRLab.htm
***********************************