My NT-Shims..Final Story!

Richard K. Shoemaker (rshoe@wwitch.unl.edu)
Tue, 28 Oct 1997 13:19:28 -0600

Dear All:

I'm mailing to the entire group because I've had such an outpouring of
helpful suggestions and responses. My Shims are good as new...thanks to
< $10 in components and the schematics from Acorn, Inc. What happened
is a flaky regulator and 2 dead op-amps (which had the output latched at
max).

The _surprising_ thing is that the shims are OK, even though the base of
the RT-shimset was so hot we could have fried breakfast on it!!?? All
18 shims control as they should with no apparent cross-talk...nothing
seems to have fused inside the windings.

I now have the RT shim supply plugged into a latching-relay which will
not allow things to power back up until manually reset. I already had
one of those powering my transmitters...I almost paid dearly for not
including the shims in that curcuit (I mistakenly thought I was safe
since I don't have any DACs to worry about).

Thanks again to all!

Rich S.

-- 