Hi folks,
As a follow-up from previous queries about probe power limits for when the
manufacturer's guidelines are incomplete or absent: what are people's thoughts
on the limits of fast pulsing experiments when not decoupling? I'm specifically
interested in RT probes, but if others have comments to make about helium
cryoprobes then I'm sure there will be other interested parties.
Applications for fast pulsing are for any and all fast-relaxing systems e.g.
paramagnetic systems or quadrupolar nuclei such as 11B. My specific interest
is for 19F and PFAS quantification for environmental applications, and it
seems to be common practice to use paramagnetic relaxation agents to help
with sensitivity.
The "Typical Pulses" document supplied with our Bruker RT 400 MHz BBFO iProbe
covers pulse lengths for pulse-acquire experiments at full power with D11 and
an unspecified acquisition time, which I infer to mean anything at all up to
and including AQ ~0, so I assume a limiting recycle delay of D1+AQ ~1 s.
For 1H one can go to 1 ms _at_ 8 us 90 (trim pulse), so assuming that same D1
1, would be an RF duty cycle of 0.1%. Is that D1 1 really needed, or can I
just pulse faster provided I don't exceed that RF duty cycle e.g. 100 us with
D1 0.1? (ignoring the hardcoded 30m in zg* pulse programs for now)
19F doesn't have a trim pulse specification, so you're left extrapolating from
the maximum 1000 degrees _at_ 12 us 90 (133.3 us), or a 0.013% RF duty, and an
implied minimum D1 0.09.
For 11B there is no specification, so you're extrapolating from between the
numbers for 31P and 13C. Only 13C has a trim pulse specification (2 ms _at_ 25 us 90),
but both have a 1000 degree limit _at_ 8 us 90, so 88.9 us RF per second, and an
implied minimum D1 0.09.
For CPMG its received wisdom to not pulse too much, but without there being
any clear guidelines for what "too much" is for 1H. My interest in this one
is for ligand screening in general, and specifically PEARLScreen on an RT
probe at 400 MHz where I'll take any sensitivity enhancement I can lay my
hands on.
If I take the conservative approach where one can use 1 ms of full power RF
per second, then with an 8 us 90 this corresponds to only 62 echoes. If I
take Bruker library parameter sets as being reasonable starting points, then
PROF_CPMG uses 128 echoes with an interpulse delay of 0.6 ms, albeit with D1
4 and AQ ~4 s at 400 MHz. A different CPMG-based parameter set HOS_CPMGES1D
uses 232 echoes for the same D1+AQ and a shorter interpulse delay of 0.2 ms,
so yet more RF again. Is the longer D1+AQ what makes this OK? Or is the simple
RF duty cycle model wrong and the interrupted pulsing is what allows one to
pulse more? Or are these parameter sets unsuitable for RT probes to begin with?
Any thoughts would be appreciated - especially data-driven ones!
Thanks
Geoff
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3880): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3880__;!!PvDODwlR4mBZyAb0!WXR48JSkKph2pP1nYY0rpoO2n_6hpr=
sVe_-tqAnNfuz1I8Ng2bOglXGMl0IXpi_GThb6nuLODWP9J1t2mrlxK9K3rkeR$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121904998=
/7559972__;!!PvDODwlR4mBZyAb0!WXR48JSkKph2pP1nYY0rpoO2n_6hprsVe_-tqAnNfuz1I=
8Ng2bOglXGMl0IXpi_GThb6nuLODWP9J1t2mrlxK_mUVKts$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Oct 23 2026 - 11:01:32 MST