Many thanks for Feng Qiu and Mike's advice. The problem has been fixed by clone
the hd from another Neo 600 nearby. The machine is runing smoothly although
some warning notes for the wiring appeared because of the usage of the hd
from another spectrometer. I have checked the broken hd, there are too many
errors or disk failure to be fixed. I am not sure if the broken hd would be
fine after fixing the problems.
Best wishes
Junfeng
-----原始邮件-----
发件人:"Michael Groves via groups.io" <mgrovesnmr=yahoo.com_at_groups.io>
发送时间:2026-10-15 22:15:24 (星æœ=
Ÿ四)
收件人: "main_at_ammrl.groups.io" <main_at_ammrl.groups.io>
抄送:
主题: Re: [AMMRL] about startup failure for Bruker Neo 600 with SCU
Hi Junfeng,
I have had... multiple... customers swap out their own HD from the SCU. There
is a new drive that is being provided for those that is supposed to be more
robust than the previous ones. Swapping it out isn't too complicated. You'll
need a current nmrbackup file if possible. Here's what I recently sent to
somebody else:
First, you’ll need to swap out the drives. Make notes of what cables
are connected where on the SCU and power down the console. Disconnect the
cables and loosen the screws above and below the SCU and slide it out. The
drive is in the bottom of the board—I believe there are 4 screws in
the side of the SCU that you remove and then the drive slides out. After that,
slide the new one in. If you put the SCU upside down on a flat surface the
drive might go in easier. Once you get the drive in, put the screws in and
put the SCU back in and reconnect the cables and fire the thing up.
Once it’s fired back up and fully booted you’ll need to update
spectros. Open a terminal and run the following command:
“/opt/Bruker/spectros/topspin/python/bin/python3 /opt/Bruker/spectros/post/updatespectros.py”
and it should run the updater and then reboot the rack. Once the rack has rebooted, run cf.
After a successful cf, you’ll need to run nmr_restore and restore the newest backup. And
then cf again. And then you can do edprobe and expinstall. After that you should be back up and running.
One note--the updatespectros command depends on what version topspin and OS
you have installed. If you open a terminal or a command prompt and cd to
either C:\Bruker\spectros\post or /opt/Bruker/spectros/post you can see what
the script is. If you're running Linux you might just have updatespectros.sh
which you can just run with the command "sh updatespectros.sh" If you've got
Windows then most likely you have a python script in which case you can use
the command:
C:\Bruker\spectros\topspin\python\python C:\Bruker\spectros\post\updatespectros.py
And in some weird cases I think I've also seen a perl script. I believe the
command is the same you just have to point it at perl instead of python.
One other way to update spectros on the new HD is to run the topspin installer
if you've got it. Just click through the defaults and when it asks you what
kind of installation you want to do, select customized and then uncheck
everything except spectros. This will do the equivalent of the above commands
and is arguably simpler.
And then as stated above, after you get the right version of spectros installed
to the drive, run cf and then nmr_restore, and then cf again, and then edprobe.
After that you should be back up and running.
We have seen multiple failures of the original hard drives and sometimes people
want a service visit to swap it out but there have been plenty who have done
it on their own without issues.
Cheers,
Mike
On Thursday, October 15, 2026 at 06:46:21 AM MDT, 向俊锋 <jfxiang_at_iccas.ac.cn> wrote:
Dear spinners
Yesterday, we failed to startup the bsms unit with blue light flickered on
aqs unit with Bruker Neo 600 nmr spectrometer. According to the suggestion
nearby, it is suspected that the failure of hard disks in scu unit. I just
wonder if anyone has experience in change the hard disk from scu. Could we
handle it by ourselves? I do really appreciate if anyone share the experience.
Best wishes
Junfeng
ICCAS, Beijing China
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3873): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3873__;!!PvDODwlR4mBZyAb0!TM2MjrwfzmeHtnNWU6vqYQSc1EyS1X=
aG_tYzMSurNYZe0k3Gdw4CLzvFrDn3w4yZlfkoVJmAFxtsu9HLHA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121768988=
/7559972__;!!PvDODwlR4mBZyAb0!TM2MjrwfzmeHtnNWU6vqYQSc1EyS1XaG_tYzMSurNYZe0=
k3Gdw4CLzvFrDn3w4yZlfkoVJmAFxuc4qxbFA$
awaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Oct 16 2026 - 06:03:07 MST