Re: [AMMRL] about startup failure for Bruker Neo 600 with SCU

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Thu, 15 Oct 2026 14:15:24 +0000 (UTC)

 Hi Junfeng,

I have had...  multiple... customers swap out their own HD from the SCU. 
There is a new drive that is being provided for those that is supposed to be more
robust than the previous ones.  Swapping it out isn't too complicated. 
You'll need a current nmrbackup file if possible.  Here's what I recently
sent to somebody else:

First, you’ll need to swapout the drives.  Make notes of what
cables are connected where on the SCUand power down the console.  Disconnect
the cables and loosen the screwsabove and below the SCU and slide it out. 
The drive is in the bottom ofthe board—I believe there are 4 screws in
the side of the SCU that you removeand then the drive slides out.  After
that, slide the new one in.  Ifyou put the SCU upside down on a flat surface
the drive might go ineasier.  Once you get the drive in, put the screws in
and put the SCU backin and reconnect the cables and fire the thing up.

 

Once it’s fired back up andfully booted you’ll need to update
spectros.  Open a terminal and run thefollowing command:
“/opt/Bruker/spectros/topspin/python/bin/python3/opt/Bruker/spectros/post/updatespectros.py”
and it should run the updater andthen reboot the rack.  Once the rack has rebooted, run cf. 
After asuccessful cf, you’ll need to run nmr_restore and restore the newestbackup.  And
then cf again.  And then you can do edprobe and expinstall.  After that you should be back
up and running.

One note--the updatespectros command depends on what version topspin and OS
you have installed.  If you open a terminal or a command prompt and cd
to either C:\Bruker\spectros\post or /opt/Bruker/spectros/post you can see
what the script is.  If you're running Linux you might just have updatespectros.sh
which you can just run with the command "sh updatespectros.sh"  If you've
got Windows then most likely you have a python script in which case you can
use the command:
C:\Bruker\spectros\topspin\python\python C:\Bruker\spectros\post\updatespectros.py 
And in some weird cases I think I've also seen a perl script.  I believe the
command is the same you just have to point it at perl instead of python.  
One other way to update spectros on the new HD is to run the topspin installer if
you've got it.  Just click through the defaults and when it asks you what kind
of installation you want to do, select customized and then uncheck everything except
spectros.  This will do the equivalent of the above commands and is arguably simpler.  
And then as stated above, after you get the right version of spectros installed to
the drive, run cf and then nmr_restore, and then cf again, and then edprobe. 
After that you should be back up and running.

We have seen multiple failures of the original hard drives and sometimes people
want a service visit to swap it out but there have been plenty who have done it on
their own without issues.  

Cheers,Mike


 On Thursday, October 15, 2026 at 06:46:21 AM MDT, 向俊é8B <jfxiang_at_iccas.ac.cn> wrote:

Dear spinners

Yesterday, we failed to startup the bsms unit with blue light flickered on
aqs unit with Bruker Neo 600 nmr spectrometer. According to the suggestion
nearby, it is suspected that the failure of hard disks in scu unit. I
just wonder if anyone has experience in change the hard disk from scu. Could
we handle it by ourselves? I do really appreciate if anyone share the experience.

 Best wishes

 Junfeng

 ICCAS, Beijing China


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3866): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3866__;!!PvDODwlR4mBZyAb0!S0Vx3qPOd2_Ru68U5rlHegzN9J2Iss=
I67KqYRhiZimhaLKpcbTgbixR_NBgcwCBkinaQlQNzWUkOyTEwLQ_2R0sAWIRB$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121768988=
/7559972__;!!PvDODwlR4mBZyAb0!S0Vx3qPOd2_Ru68U5rlHegzN9J2IssI67KqYRhiZimhaL=
KpcbTgbixR_NBgcwCBkinaQlQNzWUkOyTEwLQ_2R8mOtmnn$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Oct 15 2026 - 08:01:27 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST