Hi Ken,
Just a simple mix of ethylene glycol and house generated R.O. water. Despite
being closed loop the volume does gradually dissipate. Every few months I
refill with 1:4 ratio to reliably keep it above 15% glycol. The chiller is
long in the tooth and afaik has never had a problem with algae growth.
As an aside, I would intuit algae contributing towards wear-and-tear for the
pump and motor. We get about 2 reliable years out of a rotary vane pump without
algae. Keeping the heat exchanger clean of dust and debris has weekly/monthly \
maintenance checks, and if it's on building exterior then seasonal pollen,
leaves, and snow become major considerations.
Kind regards,
Luke
Luke Fulton, PhD
CHEM BLDG R003
NMR Core Facility Director
Unit 3060
COR2E & Department of Chemistry
55 N Eagleville Road
University of Connecticut
Storrs, CT 06279
________________________________
> From: main_at_ammrl.groups.io on behalf of Kenneth Sharp-Knott via groups.io
> Sent: Monday, October 12, 2026 11:19 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] Water chiller biocide/algaecide recommendations for NMR & diffusion systems
Dear AMMRL,
What biocide or algaecide do you recommend for closed-loop water chillers
supporting NMR systems (diffusion gradient amplifiers and probe recirculator s)?
We want to prevent biological growth without causing foaming or corroding
iternal copper, brass, and stainless-steel components. Specific product names
and dosage routines that work well in your facilities would be greatly appreciated.
Thanks,
Ken Sharp-Knott
Senior Manager, NMR & Analytical Core Facilities
Department of Chemistry
Virginia Tech
(540)267-6502 (Cell)
(540)231-0885 (Office)
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3851): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3851__;!!PvDODwlR4mBZyAb0!RQktlpx_A_nFAaSoZRE7rvkfdzloO0=
F0L7QiIbpISfaAyWj-JOFPWdcocRconNDVQmd_8xssaCvGlk1QYriMaxLsHHJN$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121716773=
/7559972__;!!PvDODwlR4mBZyAb0!RQktlpx_A_nFAaSoZRE7rvkfdzloO0F0L7QiIbpISfaAy=
Wj-JOFPWdcocRconNDVQmd_8xssaCvGlk1QYriMa0Cmyrsi$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Mon Oct 12 2026 - 10:01:44 MST