Re: [AMMRL] Water chiller biocide/algaecide recommendations for NMR & diffusion systems

From: Albers, Thomas via groups.io <talbers=nmhu.edu_at_groups.io>
Date: Mon, 12 Oct 2026 15:30:56 +0000

Our x-ray diffractometer uses Polyscience Polyclear Mix 30 algaecide/corrosion
inhibitor by recommendation from the manufacturer (Rigaku). If anyone wants
to mix their own, that's Polyquat 30 algaecide combined with benzotriazole
(1000 ppm IIRC) adjusted to basic pH to minimize corrosion.

Thomas

________________________________
> From: main_at_ammrl.groups.io on behalf of Kenneth Sharp-Knott via groups.io
> Sent: Monday, October 12, 2026 09:19
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] Water chiller biocide/algaecide recommendations for NMR & diffusion systems


Dear AMMRL,

What biocide or algaecide do you recommend for closed-loop water chillers
supporting NMR systems (diffusion gradient amplifiers and probe recirculators)?

We want to prevent biological growth without causing foaming or corroding
internal copper, brass, and stainless-steel components. Specific product
names and dosage routines that work well in your facilities would be greatly
appreciated.

Thanks,

Ken Sharp-Knott
Senior Manager, NMR & Analytical Core Facilities
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3845): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3845__;!!PvDODwlR4mBZyAb0!WwSEF3_FR8-vaOpUticiDXS2UPGU6T=
Xkdlly14C1T2pRl_zuAUEOInZ_WMI1gFMlFue_Kyo8GY-6wXQ2MpAH708lghk9$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121716773=
/7559972__;!!PvDODwlR4mBZyAb0!WwSEF3_FR8-vaOpUticiDXS2UPGU6TXkdlly14C1T2pRl=
_zuAUEOInZ_WMI1gFMlFue_Kyo8GY-6wXQ2MpAH73nOcJuq$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Oct 12 2026 - 08:46:39 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:11 MST