Re: [AMMRL] young tube broken inside the magnet

From: Ulrich via groups.io <Ulrich.Haunz=uni-konstanz.de_at_groups.io>
Date: Thu, 08 Oct 2026 16:26:13 +0200

Dear all!

Am Donnerstag, Oktober 08, 2026 02:45 CEST, schrieb "Blumenfeld, Alexander via groups.io:

> Hello All,
>
> The Young tube got broken on the way out from the magnet. I was able to get
> the broken top of the tube, but the rest (the tube plus the spinner) got
> stuck inside. The emergency red button does not bring the tube out; neither
> does the "lift" BSMS button. Is there any other way to solve the problem
> other than taking the probe out and pull the spinner out? Any suggestion
> is appreciated!
>

This is a good chance to get the probe out of the magnet. I suggest this is
a Bruker NMR.
Please mark the direction of the shim system inside the magnet. There are
three nice and long screws from the top to the inner part of the shim system.
These has to be srewed out - and also you have to loosen the shim system
at the bottom clamping device, too.

Then you can get the glas tube in the spinner out of your system. Is it the
JEOL one - removing the probe is all you can do here. Then you can get the
sample in the rotor out of the bottom side. Sorry - this is the work you
can do now.

Von welcher Uni stammst Du? ... Blumenfeld ist ein deutscher Name.

Yours sincerely,

Ulrich Haunz

> Alex Blumenfeld
>

--
=========================

Ulrich Haunz, Dipl.-Ing. (FH)
NMR-Spektroskopie

Raum L 710, L 807
NMR-Center M520, M521, M527
Postfach M 715
Fachbereich Chemie
Universitaet Konstanz

78457 Konstanz

Tel.: +49 (0)7531 88-4261 -> M521 Bruker Avance NEO 800
Tel.: +49 (0)7531 88-4422 -> L807 JEOL Lambda 400
Tel.: +49 (0)7531 88-4955 -> L710 Bruker AVIII HD 400
Tel.: +49 (0)7531 88-4956 -> M520a Bruker AVIII 600
Tel.: +49 (0)7531 88-4957 -> M527 Bruker AVIII 400 Automation
Tel.: +49 (0)7531 88-4958 -> M520 Bruker AVIII 400 WB
Tel.: +49 (0)170 9153213
FAX : +49 (0)7531 88-3898
=========================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3834): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3834__;!!PvDODwlR4mBZyAb0!UjmajTr593e9BzdmUU3WyPbCt8MHUO=
31Qh8o08_5xXouJPVynwAXTNo4cb2y-iCtx6wITsl8nzviDhkvkIOkkywP-gRc$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121655142=
/7559972__;!!PvDODwlR4mBZyAb0!UjmajTr593e9BzdmUU3WyPbCt8MHUO31Qh8o08_5xXouJ=
PVynwAXTNo4cb2y-iCtx6wITsl8nzviDhkvkIOkkyKY_P1j$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-

Received on Thu Oct 08 2026 - 07:57:10 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:11 MST