Re: [AMMRL] young tube broken inside the magnet

From: Fulton, Luke via groups.io <fko24003=uconn.edu_at_groups.io>
Date: Thu, 8 Oct 2026 16:47:23 +0000

Hi Alex,

Any time a tube breaks inside the magnet it's generally wise to drop the probe.
I always do that before mounting a rescue.
-can't know for certain full sample condition just by observing from above
the bore opening
-tube breaks below the spinner will leave glass pieces behind in/on the probe/shim
stack even after spinner is retrieved
-removing glass shards and cleaning spilled sample contents from probe can't/shouldn't
be done inside the magnet
-"fishing" out the sample sometimes goes awry, if it falls back down freely
it can impact and risk further damage
-possible to push sample out from below if probe is removed

A second pair of hands goes a long way as well. Best of luck!

Kind regards,
Luke

Luke Fulton, PhD
CHEM BLDG R003
NMR Core Facility Director
Unit 3060
COR2E & Department of Chemistry
55 N Eagleville Road
University of Connecticut
Storrs, CT 06279


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3837): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3837__;!!PvDODwlR4mBZyAb0!WSQeauVh_rzBblNhV15gDHp4szQpxH=
N1IaaAj-P-Wx6NT59oihi1it9C_LGd360UBzGpZjklHjbc7UyyEE4dpd4ctLAY$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121655142=
/7559972__;!!PvDODwlR4mBZyAb0!WSQeauVh_rzBblNhV15gDHp4szQpxHN1IaaAj-P-Wx6NT=
59oihi1it9C_LGd360UBzGpZjklHjbc7UyyEE4dpVn7kZ0v$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Oct 08 2026 - 10:04:41 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:11 MST