Hi,
I could be way off the mark on this one - but if wobb does not work
either, then your problem may be completely unrelated to ATM. If the
wobb command does not initiate the expected software response (the wobb
experiment window), then its possible you are dealing with a problem
with Topspin, console or preamp instead. If wobb starts normally, but
does not lead to a recognisable wobb curve (no dip visible), then its
quite possible there is a probe problem, but it may or may not related
the ATM unit.
Indrek
--
Indrek Reile, Ph.D.
Senior Research Fellow / Solution NMR
Member of Science Council
National Institute of Chemical Physics & Biophysics
Akadeemia tee 23, 12618 Tallinn, Estonia
email:indrek.reile_at_kbfi.ee
mobile: +372 523 6054
23.10.2025 20:34 Baleja, James D via groups.io kirjutas:
> Thanks all for your useful insights - even found the wrench to help
> dislodge the flush air tubing.
>
> However the gradient controller is still stuck - will try again
> Monday. Bit easy enough to take the probe out with only one cable
> attached.
>
> Wobb and Atmm still don't work, so I think the next step is to send
> the probe out for repairs
>
> Jim
>
>
>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2954): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2954__;!!PvDODwlR4mBZyAb0!VqdBTtqb7D7muV6A-dtUYr59Egk6hJ=
pZSYFG5aytIJVWuzwQ5yDpxKH4ceKyg45_wDRVFbVZUkL3d0TE-hN3BcX02hKE$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115912971=
/7559972__;!!PvDODwlR4mBZyAb0!VqdBTtqb7D7muV6A-dtUYr59Egk6hJpZSYFG5aytIJVWu=
zwQ5yDpxKH4ceKyg45_wDRVFbVZUkL3d0TE-hN3BRJddTfR$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Oct 24 2025 - 06:00:27 MST