From: Eugenio Alvarado via groups.io <ealva=umich.edu_at_groups.io>
Date: Tue, 21 Oct 2025 15:19:36 -0400

Hi all,

Our Varian VnmrS 700 has developed a problem where all the LEDs of all
the digital controllers are off except for PSOK. I can't communicate
with anything except for the ProTune unit.  All the controllers show a
link LED to the network switch. I swapped the network switch but this
made no difference. The power supply appears to be ok as I checked the
voltages from the power supply and they are all 0.1-0.2 V above their
nominal values except for the 24 V output which is at 24.0 V.

If anyone has any idea or suggestion to try, I would greatly appreciate it.

Thanks,
Eugenio

--
Eugenio Alvarado, Ph.D.
NMR Facility Director
Chemistry Department, room 3500. University of Michigan
930 N. University Ave., Ann Arbor, MI 48109-1055.
Voice: (734)-764-9924


-=-=-=-=-=-=-=-=-=-=-=-