Re: [AMMRL] BOSS matrix loading problem

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Thu, 31 Oct 2024 01:18:59 +0000 (UTC)

Hi Gopal,

The most common cause for this is that you swapped the shim cables.  I
believe the A connector in the shim system has a couple of resistors that
are used by the hardware to identify which shim system you've got, and if
you swap the cables the correct resistances aren't found and the shim system
ID is nonsense and the instrument gets mad.  

You can also go into the service webpage for the BSMS (use the command ha in
topspin, then click Shim and I believe Shim Configuration after that) and see
what it thinks that the shim system ID is.  Sometimes the resistors drift
a bit and the shim system ID just changes from what it should be by a little bit. 
If the shim system is close to what it should be (302928 instead of 302927 for
example) you can put the part number of the shim coil in as a substitute identifier. 
I would only do this if I had checked that the cables weren't swapped.  If they
are swapped, that number comes up as nonsense and you shouldn't use this substitute.

Cheers,Mike

On Wednesday, October 30, 2024 at 02:18:54 PM MDT, Gopal Subramaniam wrote:

Dear Spinlanders,

We have AvanceIII Z gradient with topshim and I ran into problems after a cleanup
of lower shim stack to correct sample spin problem. I had to disconnect and
re-connect shim cables during the process.

Now,the system is giving an error that BOSS matrix file could not be loaded. 
Error details say 'failed due to BOSS file and HW code of connected shim system do
not match'.

I tried to go to web interface and tried to manually load the .dat file that was
in use before. Even though no parsing error shows, the file doesn't appear. 
Without this all shim currents are 0 and the shims do not show up at all in the
bsms.

Hope you can give me some guidance on fixing this. 

Thank You.

Dr. Gopal Subramaniam (X74123)

Department of Chemistry and Biochemistry
Queens College, CUNY
Queens, NY 11367
 


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1802): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1802__;!!PvDODwlR4mBZyAb0!TVc93eFk6t_FggBPa0wjqpf-gpghAr=
g2hdPHQB8eN7KonmnkEsRy2yxH-7E1To6t9cXKhEdd_s3TtFFVah1XPo0eOWuK$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109303421=
/7559972__;!!PvDODwlR4mBZyAb0!TVc93eFk6t_FggBPa0wjqpf-gpghArg2hdPHQB8eN7Kon=
mnkEsRy2yxH-7E1To6t9cXKhEdd_s3TtFFVah1XPj72jfiO$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Oct 30 2024 - 18:19:17 MST

This archive was generated by hypermail 2.4.0 : Fri Nov 01 2024 - 17:08:12 MST