Hi Jun,
To echo others comments, it works well to use the 'gas' supply of a high
pressure liquid nitrogen tank. We had that set up at my previous job to
supply bearing/drive gas for SSNMR. We also had a nice setup where you could
connect two HP liquid nitrogen tanks and the flow automatically switched
when one tank went empty. That eliminates any disruption in flow and allows
you to switch in another tank while maintaining flow.
Good luck,
Vernon
Vernon Russell
Analytical Team / Gore Incubator
Enterprise Analytical Team
https://wlgore.sharepoint.com/sites/Intranet-EST/SitePages/CT-Analytical-Team-Strategy-and-Organization.aspx
vrussell_at_wlgore.com
T +1-410-506-3852
W. L. Gore & Associates, Inc.
Together, improving life
501 Vieve's Way, P.O. Box 1320
Elkton, MD 21922-1320
USA
gore.com
________________________________
> From: main_at_ammrl.groups.io on behalf of Gu, Jun <jungu_at_sas.upenn.edu>
> Sent: Wednesday, October 16, 2024 1:53 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] high pressure liquid nitrogen tank to support prodigy probe during house nitrogen shut down
Hello All,
There will be a house nitrogen shutdown for 3 day in my building. To keep the
NMRs available for research, I am considering using high-pressure liquid
nitrogen tanks to supply three spectrometers with SampleCase: (1)NMR with
a IProbe and BCU I, (2) NMR with a Prodigy probe and BCU I, (3) NMR with a
Helium-cooled probe (no SampleCase).
Is this approach feasible? How many liquid nitrogen tank should I order?
I would greatly appreciate it if anyone could share their experience or advice.
Best,
Jun
This email may contain trade secrets or privileged, undisclosed or otherwise
confidential information. If you have received this email in error, you are
hereby notified that any review, copying or distribution of it is strictly
prohibited. Please inform us immediately and destroy the original transmittal.
Thank you for your cooperation.
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1754): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1754__;!!PvDODwlR4mBZyAb0!S2-aczO4nVP-vleKMwP86QibGvgroj=
ESnY9ByXT6g-uMluZTGnYKFxVGvzG4kN3kElc_SYdwA2oO4aPf$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109048916=
/7559972__;!!PvDODwlR4mBZyAb0!S2-aczO4nVP-vleKMwP86QibGvgrojESnY9ByXT6g-uMl=
uZTGnYKFxVGvzG4kN3kElc_SYdwA47UijYT$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Oct 16 2024 - 11:16:08 MST