Re: [AMMRL] high pressure liquid nitrogen tank to support prodigy probe during house nitrogen shut down

From: Richard Fitch via groups.io <richard.fitch=indstate.edu_at_groups.io>
Date: Wed, 16 Oct 2024 18:10:15 +0000

Hi Jun,

I would recommend using the boiloff from a liquid nitrogen dewar. It will
last longer and you can avoid using magnetically attractive cylinders.

Cheers,
Rick

Richard W. Fitch
Professor, Department of Chemistry and Physics
Councilor, Wabash Valley Section, American Chemical Society
Indiana State University
600 Chestnut Street
Science Building, Rm. S35E
Terre Haute, IN 47809
Phone: (812) 237-2244
Fax: (812) 237-2232

"Do an apparently simple experiment that gives you an important bit of information..."
Julius Axelrod

The love of money is the root of some evil. Fear and insecurity are the root of most of it.

This email, and any attachments, thereto, is intended only for use by the
addressee(s) named herein and may contain privileged and/or confidential
information. If you are not the intended recipient of this email, you are
hereby notified that any dissemination, distribution or copying of this email,
and any attachments thereto, is strictly prohibited.

> From: main_at_ammrl.groups.io On Behalf Of Sham, Shing via groups.io
> Sent: Wednesday, October 16, 2024 2:07 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] high pressure liquid nitrogen tank to support prodigy probe during house nitrogen shut down

Hi Jun,

I did it once before. Depending on the usage, those tanks can run out pretty
fast. I would suggest to put a T to connect two tanks. Just swap it when
one is running low.

Good luck.

Simon

> From: main_at_ammrl.groups.io On Behalf Of Gu, Jun via groups.io
> Sent: Wednesday, October 16, 2024 12:54 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] high pressure liquid nitrogen tank to support prodigy probe during house nitrogen shut down

Hello All,

There will be a house nitrogen shutdown for 3 day in my building. To keep the
NMRs available for research, I am considering using high-pressure liquid
nitrogen tanks to supply three spectrometers with SampleCase: (1)NMR with a
IProbe and BCU I, (2) NMR with a Prodigy probe and BCU I, (3) NMR with a
Helium-cooled probe (no SampleCase).

Is this approach feasible? How many liquid nitrogen tank should I order?
I would greatly appreciate it if anyone could share their experience or advice.


Best,

Jun



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1753): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1753__;!!PvDODwlR4mBZyAb0!WGxixJSojL9AEYPb9AQRrjcn2KOFtM=
YOaq--jsdIHBVutKwcPjisGdODUqg3vwxkTiswtLBMGHoEiZ9aGmQdZAkgEUOU$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109048916=
/7559972__;!!PvDODwlR4mBZyAb0!WGxixJSojL9AEYPb9AQRrjcn2KOFtMYOaq--jsdIHBVut=
KwcPjisGdODUqg3vwxkTiswtLBMGHoEiZ9aGmQdZFUyuPCU$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Oct 16 2024 - 11:10:23 MST

This archive was generated by hypermail 2.4.0 : Fri Nov 01 2024 - 17:08:12 MST