Re: [AMMRL] AMMRL: Bruker MAS probe rotor ejection problem

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Mon, 7 Oct 2024 15:40:47 +0000 (UTC)

 Hi Weiguo,

When you click the eject button does the air turn on and the sample not come
out, or does the air not turn on?   It does sound like your probe
could use some TLC.  It's possible that something has sprung a leak. 
It's also possible that the top of the stator is not aligned well with the
bottom of the transfer tube and that needs to be adjusted.  

Cheers,Mike

    On Monday, October 7, 2024 at 08:28:57 AM MDT, weiguo.hu_at_mail.pse.umass.edu wrote:

Dear Colleagues:

We have a 4mm CP/MAS VTN probe that has a problem. We usually set the EJ gas to
ca. 30% and the ejection usually works well and the rotor comes out with not too
much of a force to bang against the rod ceiling. In recent months, however,
occasionally the ejection action would not activate upon clicking the eject
button on the MAS control. I would need to increase the eject pressure to ca.
60% to get the ejection going again, then lower it back to 30% and it will
behave for a while. However, the ejection gas pressure needed to re-activate the
action has been increasing over time and now even with eject gas at nearly 100%
the ejection action would not kick in. I suspect that one of the gas tubings in
the probe might be aging and losing flexibility which might have caused this
problem. I wonder if anybody has done some research about this problem and
can share some wisdom.

Thank you,

Weiguo Hu
Director of NMR Labs/Chemistry & Materials
Conte Polymer Research Center
120 Governors Dr
University of Massachusetts
Amherst, MA 01003


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1722): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1722__;!!PvDODwlR4mBZyAb0!XWsv5mw_gggf_a4yeNC5gUZRzC0edq=
DqXTZhXgx6j3_BSuMpuV2klojmQHCXGWNCByweXATSSsok6bQvb8Ti4Fh9lcXM$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108868672=
/7559972__;!!PvDODwlR4mBZyAb0!XWsv5mw_gggf_a4yeNC5gUZRzC0edqDqXTZhXgx6j3_BS=
uMpuV2klojmQHCXGWNCByweXATSSsok6bQvb8Ti4Hy0SAaq$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Oct 07 2024 - 08:40:57 MST

This archive was generated by hypermail 2.4.0 : Fri Nov 01 2024 - 17:08:12 MST