Re: [AMMRL] NMR scripting idea - seeking feasibility opinions...

From: Brian Antalek via groups.io <bantalek=me.com_at_groups.io>
Date: Tue, 25 Nov 2025 10:28:23 -0500
have a session with ChatGPT to create a py script. 

-Brian
cosmic message from myPhone

On Nov 25, 2025, at 10:25 AM, Akien, Geoffrey via groups.io <g.akien=lancaster.ac.uk_at_groups.io> wrote:

Hi Ken,

 

I can see the motivation for this, but I think Garbage In Garbage Out applies for this one, in the sense that you can’t practically control for sample quality. You’d have to have a way of ignoring results when your solvent peak isn’t resolved, proper handling of multiplets and JHD couplings, and then have a way of handling the many false positive when the lineshape is poor because users have floating solids, incomplete mixing, scratched tubes etc. etc.

 

True, you might be able to detect gross changes from cracked inserts or something like that, but usually you’ll hear from users about that one pretty fast as it should be obvious even for non-experts.

 

Degradation of 3D shims is going to be more of a gradual process, and I’d question whether you’d be able to reliably pull that out of the noise. The most robust data is always going to be from a validated and reproducible test sample e.g. sucrose, which is presumably why the Bruker AutoCalibrate system relies on a particular version of this.

 

Thanks


Geoff

 

From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Kenneth Sharp-Knott via groups.io
Sent: 25 November 2025 15:06
To: main_at_ammrl.groups.io
Subject: [External] [AMMRL] NMR scripting idea - seeking feasibility opinions...

 

Hello everyone, I've brainstormed an idea and wondering if we have reached the point were technology might make this a possibility.

 

The VT NMR facility uses LOGS as our data management solution which is integrated with MNova via MHub.  Now that MNova and LOGS both have a Python suite, I wonder if a 'shim maintenance script' might be feasible.

 

Essentially, LOGS already knows the selected solventand with MHub, MNova could screen each sample, identify the solvent peak, and check the lineshape.  When it observes tailing, splitting, etc, it could send an email or possibly put a flag in LOGS alerting the facility staff to check the probe for contamination or a touch up on the 3d shims.

 

Curious if anyone has thoughts on this.

 

Thanks

 

Ken Sharp-Knott

Manager of Analytical Services and the NMR Facility

Department of Chemistry 

Virginia Tech

 

(540)267-6502 (Cell)

(540)231-0885 (Office)

_._,_._,_

Groups.io Links:

You receive all messages sent to this group.

View/Repl= y Online (#3003) | Reply to Group | Reply to Sender | Mute This To= pic | New Topic
Your Subscription | Contact Group Owner |

_._,_._,_
Received on Tue Nov 25 2025 - 08:16:26 MST

This archive was generated by hypermail 2.4.0 : Mon Dec 01 2025 - 17:39:38 MST