[AMMRL] NMR scripting idea - seeking feasibility opinions...

From: Kenneth Sharp-Knott via groups.io <kknott=vt.edu_at_groups.io>
Date: Tue, 25 Nov 2025 15:05:51 +0000

Hello everyone, I've brainstormed an idea and wondering if we have reached
the point were technology might make this a possibility.

The VT NMR facility uses LOGS as our data management solution which is
integrated with MNova via MHub. Now that MNova and LOGS both have a Python
suite, I wonder if a 'shim maintenance script' might be feasible.

Essentially, LOGS already knows the selected solvent and with MHub, MNova
could screen each sample, identify the solvent peak, and check the lineshape.
When it observes tailing, splitting, etc, it could send an email or possibly
put a flag in LOGS alerting the facility staff to check the probe for
contamination or a touch up on the 3d shims.

Curious if anyone has thoughts on this.

Thanks

Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3001): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3001__;!!PvDODwlR4mBZyAb0!VDvnWx0v8_lgo_nu-6hRROu4MqXtPR=
kd8LdLD2SAViXiqazvB1O_cqABBWCu3Vl7XJnZ8dNs0sFgc-zy1rg_ZTxn$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/116469977=
/7559972__;!!PvDODwlR4mBZyAb0!VDvnWx0v8_lgo_nu-6hRROu4MqXtPRkd8LdLD2SAViXiq=
azvB1O_cqABBWCu3Vl7XJnZ8dNs0sFgc-zy1pSznLvp$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Nov 25 2025 - 07:08:48 MST

This archive was generated by hypermail 2.4.0 : Mon Dec 01 2025 - 17:39:38 MST