Re: [AMMRL] Proton with Selective 31P Decoupling

From: Steve Clemens via groups.io <steven.clemens=novachem.com_at_groups.io>
Date: Tue, 04 Nov 2025 08:09:32 -0800

Ken,

Not with selectively decoupling a particular 31P nucleus specifically, but
in my experience a 1H-31P HMBC works VERY well.  I sometimes observe
couplings up to 5 bonds distant.

hmbcgplpndqf, GPZ1=60, GPZ2=45, GPZ3=57.6, all SMSQ10.100 shape


-=-=-=-=-=-=-=-=-=-=-=-