Re: [AMMRL] AMMRL: MAS III Unit on Older AVANCE III Consoles

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Thu, 21 Nov 2024 15:07:34 +0000 (UTC)

 Hi Pat,

The MAS/3 is directly compatible with an AVIII console--it just connects to
the ethernet switch in the console and as long as you've got topspin version
3.5 or newer it will integrate seamlessly.  No need to do anything special,
and you can run it through the masdisp command in topspin.

If you do have an older console or version of topspin you can operate it through
the webpages.  Basically you just have to set up the ethernet connection on
the PC to use the right IP address and then you can connect to the MAS/3 directly
through an ethernet browser.  

Cheers,Mike

On Thursday, November 21, 2024 at 07:31:34 AM MST, Patrick Szell via groups.io wrote:

Hi all,

Has anyone used Bruker’s new MAS III unit on an older AVANCE III
console with an older probe?   Apparently we would need to control
the MAS III unit through a webpage, as our console does not support MAS III
through TopSpin.

Unfortunately, our MAS I unit is beginning to have some issues, and Bruker
no longer has these in stock.


Thank you,


Pat

--

Dr. Patrick M.J. Szell

Gestionnaire du laboratoire de RMN | NMR Facility Manager
Université d'Ottawa | University of Ottawa
Faculté des sciences | Faculty of Science
E: pszell_at_uottawa.ca
T: 613-562-5800 ext 6077


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1880): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1880__;!!PvDODwlR4mBZyAb0!WkRTPR7Jd57grjKTrYWcZp62NKCpsS=
cgVOqdyIbQBMXBCTXhY6Sd000vUFpQkHzXT8jmhXMVgN0Y8S8pO5jJoBm7ovF5$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109704268=
/7559972__;!!PvDODwlR4mBZyAb0!WkRTPR7Jd57grjKTrYWcZp62NKCpsScgVOqdyIbQBMXBC=
TXhY6Sd000vUFpQkHzXT8jmhXMVgN0Y8S8pO5jJoJBQwi3l$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Nov 21 2024 - 07:07:41 MST

This archive was generated by hypermail 2.4.0 : Sun Dec 01 2024 - 13:25:02 MST