Regulator at the samplexpress should be reading 4 bar or so. At the
wall you should have something higher than that.
Cheers,Mike
On Wednesday, November 20, 2024 at 01:57:32 PM MST, Beth Moscato via groups.io wrote:
Hi all –
We have a Sample Express Lite (2016, if that matters). We’ve had
issues with the house air and related tubing recently, which has led to a lot
of adjustments to the system.
What is the recommended pressure of air to put into the Sample Xpress Lite?
Specifically, the pressure for the back dial?
Thanks!
Beth
-=-=-=-=-=-=-=-=-=-=-=-