From: Matthew Devany via groups.io <mdevany=hunter.cuny.edu_at_groups.io>
Date: Sat, 9 May 2026 09:34:55 -0400

Purush,

What you see there is background from the probe(solid teflon or other
PFAS). Try to use narrower windows, avoid that section of the shift
range if possible, and correct baseline with the macro abs.13c.

Matthew


On 5/8/26 16:59, pothula purushotham reddy via groups.io wrote:
>
>
> Hi there,
>
> I am measuring *¹⁹F NMR* on a *Prodigy probe* at *400 MHz*, but the
> spectra look strange/weird. Does anyone know what might be causing
> this or how to fix it? Any suggestions would be greatly appreciated.
>
> Thank you!
>
> Best
> Purush
>
>
>
--
*Matthew Devany, Ph.D.*
Director NMR Laboratory -- Department of Chemistry, Hunter College of
the City University of New York
695 Park Ave., Rm. N1302 -- New York, NY 10065
Lab: 212-772-5337 -- Fax: 212-396-6327 --
https://urldefense.com/v3/__https://hunter.cuny.edu/artsci/chemistry/resear=
ch/nuclear-magnetic-resonance/__;!!PvDODwlR4mBZyAb0!XRcHTO3RNzDYQVgrI6ZUGa1=
OXedcmx18d0d-SAag0w5ZDraUU8kjzSb-38dX7j_iEnJqXYF8i3AQPtUPXahNCosLdY_b$



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3438): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3438__;!!PvDODwlR4mBZyAb0!XRcHTO3RNzDYQVgrI6ZUGa1OXedcmx=
18d0d-SAag0w5ZDraUU8kjzSb-38dX7j_iEnJqXYF8i3AQPtUPXahNCkL6U9AV$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/119227897=
/7559972__;!!PvDODwlR4mBZyAb0!XRcHTO3RNzDYQVgrI6ZUGa1OXedcmx18d0d-SAag0w5ZD=
raUU8kjzSb-38dX7j_iEnJqXYF8i3AQPtUPXahNCkWtlukJ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Sat May 09 2026 - 14:37:07 MST

This archive was generated by hypermail 2.4.0 : Sun May 31 2026 - 13:58:17 MST