I recently updated three NEO spectrometers, running HP-Z4 workstations, from
TopSpin 4.1.4 to 4.5.0. The results are as follows:
1. The first workstation was running CentOS 7.9. During the TopSpin software
installation three software packages were upgraded:
xcb-util-renderutil.x86_64
xcb-util-wm.x86_64
openssl11-libs.x86_64
Automatic software update was selected.
There was a problem with the firmware update for the SACA VTA that controls
the temperature of the SampleCase. Bruker Center was able to send BsmsCheckDownload_250414,
and samplecase_firmware_v13. The firmware problem was corrected successfully.
1. The next workstation was running CentOS 7.5. During the TopSpin software
installation the eight required software packages were unsuccessfully upgraded.
In addition to the three packages listed above, the five additional packages were:
fribidi.x86_64
libglvnd-glx.x86_64
libglvnd-egl.x86_64
libglvnd.x86_64
libwayland-egl.x86_64
It was not recorded which package failed to load. Since the load of the software
updates was not successful the TopSpin software update did not complete successfully.
It was decided to continue to use TopSpin 4.1.4.
1. The third workstation was running CentOS 7.9. The same packages and CheckDownload
files as the first upgrade were loaded successfully
Sincerely,
Joseph Asta
Associate Director, Engineering
AR&D, Structure Elucidation - NMR
Cell: (201) 805-8201 | joseph.asta_at_merck.com
> From: main_at_ammrl.groups.io On Behalf Of Alex Greenwood via groups.io
> Sent: Friday, May 9, 2025 12:05 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Topspin 4.5.0 in centos 7
I installed it on CentOS 7.5 a few weeks ago, despite the risks (I was hoping
it would fix my “Cannot run single acquisition” ICON bug, I
posted about that a few months ago). There was one library, libglvnd-egl.x86_64,
that failed to install, but Topspin has been running smoothly so far. ICON-NMR is
working (can’t comment on Sample Case compatibility, it works with a
Sample Xpress-lite).
Alex
> From: main_at_ammrl.groups.io On Behalf Of Nagarajan Murali via groups.io
> Sent: Friday, May 9, 2025 10:34 AM
> To: main_at_ammrl.groups.io<mailto:main_at_ammrl.groups.io>
> Subject: [AMMRL] Topspin 4.5.0 in centos 7
Dear Friends,
Any of you running Topspin4.5.0 under Centos 7 with a Neo console? If yes
let me know, please, how you managed it. Does icon nmr with sample case (24 work?
Best regards,
Nagarajan Murali
This e-mail message, together with any attachments, contains information of
Merck & Co., Inc. (126 East Lincoln Ave., P.O. Box 2000, Rahway, NJ USA 07065)
and/or its affiliates, that may be confidential, proprietary copyrighted and/or
legally privileged. (Direct contact information for affiliates is available at -
Contact us - MSD <
https://www.msd.com/contact-us/ >.) It is intended solely
for the use of the individual or entity named on this message. If you are not
the intended recipient, and have received this message in error, please notify
us immediately by reply e-mail and then delete it from your system.
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2348): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2348__;!!PvDODwlR4mBZyAb0!RS8CKidpOYc8KuZQrusiYqJ7nOJBKJ=
35uv7i1GoR16dKtAFwhf0lbJvIFpyF1xjmBebCjW_BFhgLcxsh97RZugNQFyIY$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/112900012=
/7559972__;!!PvDODwlR4mBZyAb0!RS8CKidpOYc8KuZQrusiYqJ7nOJBKJ35uv7i1GoR16dKt=
AFwhf0lbJvIFpyF1xjmBebCjW_BFhgLcxsh97RZut0mPCVG$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Mon May 19 2025 - 07:53:08 MST