I installed Topspin 4.5.0 in my Centos 7.9 host computer of my Neo 500. Ins=
talling from the iso file was smooth and I got a warning that reboot may be=
needed for license file recognition. I did a reboot and the trouble starte=
d.
The GENOME color profiles won’t start and so I was stuck with no gr=
aphical interface. We figured the nvidia drive was the issue so in /etc/x11=
we removed the line Driver “nvidia” in the file xorg.conf.=
Then it booted properly.
After starting Topspin 4.5.0 and cf everything went fine but suddenly =
Topspin reverted to data processing only mode. Mr. Michael Grover from=
Bruker Rover helped fix that issue (twice) and it seems stable as of now.=
Best,Murali
On Friday, May 9, 2025, 10:36 AM, Nagarajan Murali via groups.io <nmurali29=
=yahoo.com_at_groups.io> wrote:
Dear Friends,
Any of you running Topspin4.5.0 under Centos 7 with a Neo console? If
yes let me know, please, how you managed it. Does icon nmr with sample case
(24) work?