Re: [AMMRL] cannot reboot AVIII console

From: Hasan Celik via groups.io <hcelik=berkeley.edu_at_groups.io>
Date: Mon, 5 May 2025 09:37:39 -0700

Hi Pavel,

We encountered this error before and the problem was with the Ethernet
router that all the components were connected to. Replacing it with another
simple switch fixed it.

Best,
Hasan

_____________________________
Director, PMRC Core NMR Facility
College of Chemistry
University of California, Berkeley
227A Hildebrand Hall
Office: (510) 643-9829
Cell: (949) 732-8175


On Mon, May 5, 2025 at 9:10 AM Karel Klika via groups.io wrote:

> Dear all,
>
> We are having problems getting our console to reboot, it is getting stuck
> on "C0" at the postcode on the IPSO board when trying to reboot the
> console. The top and bottom green LEDs are lit, the middle two are not.
> Otherwise all other lights seems fine. We actually have a spare IPSO board
> and when that is inserted the postcode only reads „00” and only the top LED
> is lit. Attached are some of the error messages when TopSpin is opened, the
> first half-dozen or so from the new IPSO board and the second half-dozen or
> so from the old board. What to do to get the system working ?
>
> We are running TS3.6.2 under Windows 10 and the system is an Avance III
> 400 single bay.
>
> Any suggestions ? And thank you in advance.
>
> Yours sincerely,
> K. Klika
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2293): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2293__;!!PvDODwlR4mBZyAb0!Ut5xOTe8Qofb-CAs8IQh5T6wIJzwVv=
YLmXYJokCZ6eMgq6IM8L9Z_lAtcaHTprWEcuzbN9JxTyUHHdkq2KaOG0A70Q-n$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/112632577=
/7559972__;!!PvDODwlR4mBZyAb0!Ut5xOTe8Qofb-CAs8IQh5T6wIJzwVvYLmXYJokCZ6eMgq=
6IM8L9Z_lAtcaHTprWEcuzbN9JxTyUHHdkq2KaOG_ZX9vwI$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon May 05 2025 - 09:57:58 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 01 2025 - 15:26:23 MST