Hello Andy,
This is worrisome.
Here are my thoughts:
The account "nmrsu" is actually a computer account, not a psuedo-account
created in IconNMR using the "additional users" feature in the IconNMR
Configuration. Therefore, the password associated with the account
nmrsu should be the same as you would use when logging in to the
workstation as user = nmrsu.
If you are able to login to the nmrsu account in IconNMR without a
password, that implies that there is no password associated with that account.
Can you check that? Close TopSpin and logout of the workstation. Then
login as user: nmrsu.
If there is indeed no password required to login as nmrsu, then I
suggest that you create a password immediately.
If, however, there is already a password associated with the account
nmrsu, then I don't know why you were able to login as nmrsu in IconNMR
without a password. That seems strange to me.
NOTE: When TopSpin asks you for a password when you run commands like
"cf" or "edprosol", it is not asking for the nmrsu password, but rather
the "NMR Administrative Password." This is purely a software related
password that was created when the software was installed, and is not
associated with a specific workstation account.
I hope this helps.
Bob.
On 2024-05-24 1:50 a.m., T. Andrew Mobley via groups.io wrote:
>
>
> Hi all,
>
> We have discovered what seems to me to be an odd situation with
> respect to IconNMR and the nmrsu user (NMR Super User). We get asked
> for the nmrsu password whenever trying to change configuration of
> TopSpin (e.g. cf, edprosol, etc.) or IconNMR configuration; thus it is
> clear that nmrsu has a password and it is being used.
>
> My question comes from the following situation. In IconNMR, users are
> not allowed to stop the queue for IconNMR unless they are the user
> that started it, receiving the message to switch to the NMR Super
> User. In the past, when I received that message I just switched to my
> own account (which is all a member of the NMR Super User group), went
> through the password check and then stopped the queue. But this
> morning, I tried it with the nmrsu account. To my surprise, upon
> switching to the nmrsu within IconNMR acquisition interface (using the
> change user button), there was no password check. All of our other
> Additional users in IconNMR have passwords, and password checking is
> activated and observed for all of those accounts.
>
> I cannot find an additional switch within the IconNMR configuration,
> and unlike the Additional Users interface in IconNMR configuration,
> there doesn’t seem to be a place to set the nmrsu password within
> IconNMR.
>
> Anybody have an idea what I am missing here?
>
> Thanks for the help,
>
> Andy
>
> T. Andrew Mobley
>
> Personal Pronouns: He/Him/His
>
> For Information on Why I share my pronouns:
> http://magazine.grinnell.edu/sites/default/files/documents/Using*20Pronouns_0.pdf
>
> Virtual Office: https://grinnellcollege.webex.com/meet/mobleyt
>
> Making an Appointment: https://calendly.com/mobleyt/30min
>
> Professor of Chemistry
> Noyce Science Center 2606
> 1116 8^th Avenue
> Grinnell College
> Grinnell, IA 50112
> 641-269-4543
>
--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
---------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.
McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One Spoon"
wampum agreement.
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1284): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1284__;!!PvDODwlR4mBZyAb0!UiwMdbDJhRuNUVREBlL77kC8ztzj1P=
FpJpqJgJiqI8EAwJfYliXsXanix-CGlAIJ16a-_annnCCkgfEH$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106280679=
/7559972__;!!PvDODwlR4mBZyAb0!UiwMdbDJhRuNUVREBlL77kC8ztzj1PFpJpqJgJiqI8EAw=
JfYliXsXanix-CGlAIJ16a-_annnMeUYZoi$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri May 24 2024 - 06:22:55 MST