Hi all,

We have discovered what seems to me to be an odd situation with respect to
IconNMR and the nmrsu user (NMR Super User). We get asked for the nmrsu
password whenever trying to change configuration of TopSpin (e.g. cf,
edprosol, etc.) or IconNMR configuration; thus it is clear that nmrsu has
a password and it is being used.

My question comes from the following situation. In IconNMR, users are not
allowed to stop the queue for IconNMR unless they are the user that started
it, receiving the message to switch to the NMR Super User. In the past,
when I received that message I just switched to my own account (which is all
a member of the NMR Super User group), went through the password check and
then stopped the queue. But this morning, I tried it with the nmrsu account.
To my surprise, upon switching to the nmrsu within IconNMR acquisition
interface (using the change user button), there was no password check. All of
our other Additional users in IconNMR have passwords, and password checking
is activated and observed for all of those accounts.

I cannot find an additional switch within the IconNMR configuration, and
unlike the Additional Users interface in IconNMR configuration, there doesn’t
seem to be a place to set the nmrsu password within IconNMR.

Anybody have an idea what I am missing here?

Thanks for the help,

Andy

T. Andrew Mobley

Personal Pronouns: He/Him/His
For Information on Why I share my pronouns:
http://magazine.grinnell.edu/sites/default/files/documents/Using*20Pronouns_0.pdf

Virtual Office: https://grinnellcollege.webex.com/meet/mobleyt
Making an Appointment: https://calendly.com/mobleyt/30min

Professor of Chemistry
Noyce Science Center 2606
1116 8th Avenue
Grinnell College
Grinnell, IA 50112
641-269-4543



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1283): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1283__;!!PvDODwlR4mBZyAb0!SzyXM848Th4EvWk3in9mUA68a2cYcM=
IGLmCQORvz_Zf0hng864I5cgch_FghfnyggmESZlqvsGnH4700OTE_FBhn7sGf$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106280679=
/7559972__;!!PvDODwlR4mBZyAb0!SzyXM848Th4EvWk3in9mUA68a2cYcMIGLmCQORvz_Zf0h=
ng864I5cgch_FghfnyggmESZlqvsGnH4700OTE_FGGClDE3$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri May 24 2024 - 05:49:24 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 01 2024 - 13:37:48 MST