Hi Brian,
You might also want to ask the IVAN Spinsights community
https://ivanmr.com/,
specifically their Zulip chat. Lots of Varian specific discussion there.
To assist in troubleshooting, could you please try the following?
* In a terminal window from the host computer, run ‘rlogin master1’
* In VJ, run su.
* The terminal window should have a message with some more information about
the specific acode error.
There’s more than a few things that can cause an acode error; the
one I’m most familiar with is a strange interaction between the Z
gradient setting and the pfg control board in OpenVnmrJ3.1A, but I don’t
know a) if that’ll apply on VJ4.2, b) how that will interact with a
system with actual gradients (I run solids).
Cheers,
Alex
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1272): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1272__;!!PvDODwlR4mBZyAb0!XWAGcY4krRTkRjtxOjiUQxF1HXDRCn=
fQo3Nig4UuPzQie5rzHB4Aai5wcGU-31iu3vE3ed3vD9E6LD9cajm3gSVzPnhB$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106247663=
/7559972__;!!PvDODwlR4mBZyAb0!XWAGcY4krRTkRjtxOjiUQxF1HXDRCnfQo3Nig4UuPzQie=
5rzHB4Aai5wcGU-31iu3vE3ed3vD9E6LD9cajm3gVh-9k76$
Mute #agilent:
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mu=
tehashtag/agilent__;!!PvDODwlR4mBZyAb0!XWAGcY4krRTkRjtxOjiUQxF1HXDRCnfQo3Ni=
g4UuPzQie5rzHB4Aai5wcGU-31iu3vE3ed3vD9E6LD9cajm3gX5dJJmD$
Mute #troubleshooting:
https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!XWAGcY4krRTkRjtxOjiU=
QxF1HXDRCnfQo3Nig4UuPzQie5rzHB4Aai5wcGU-31iu3vE3ed3vD9E6LD9cajm3gdKkuoCq$
Mute #varian:
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/varian__;!!PvDODwlR4mBZyAb0!XWAGcY4krRTkRjtxOjiUQxF1HXDRCnfQo3Nig4=
UuPzQie5rzHB4Aai5wcGU-31iu3vE3ed3vD9E6LD9cajm3gcjW6x_T$
Mute #varianagilent:
https://urldefense.com/v3/__https://ammrl.groups.io/g/m=
ain/mutehashtag/varianagilent__;!!PvDODwlR4mBZyAb0!XWAGcY4krRTkRjtxOjiUQxF1=
HXDRCnfQo3Nig4UuPzQie5rzHB4Aai5wcGU-31iu3vE3ed3vD9E6LD9cajm3ge4c-lIV$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed May 22 2024 - 12:29:00 MST