Hi Luke,

With that many boards it sounds like a power issue.

Could it be a blown fuse on the AQS? The power plug at the upper right of
the back of the AQS should have a fuse (or two?) that can be checked once
the plug is removed. I personally check them with a DMM instead of by eye,
as I’m getting older…

Best,
Sameer

--
Sameer Al-Abdul-Wahid, Ph.D. | Manager
Nuclear Magnetic Resonance Centre | University of Guelph
519-824-4120 Ext. 58914 | s.wahid_at_uoguelph.ca
http://nmr.uoguelph.ca


> From: main_at_ammrl.groups.io on behalf of Fulton, Luke
> Date: Tuesday, May 14, 2024 at 15:19
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] Avance Console Trouble #Hardware #Troubleshooting

Hi Everyone,

Our Avance console had a heart attack this morning and I'm hoping the
collective knows a way to resuscitate it.

The first user of the day couldn't tune or collect spectra. There weren't
any hung processes, and the problem persisted through software and computer
restarts. Error messages pointed to multiple boards on the AQS rack (RCU,
SGU1, others) and the instrument has a history of communication issues. I
begrudgingly power cycled the console...
...and now there's no activity from a whole side of the AQS rack. CCU10, TCU3,
FCU3, GCU, RCU1/2 all dormant. pictures attached.

Any pointers would be much appreciated!
Kind regards,
Luke

Luke Fulton, PhD
CHEM BLDG R003
NMR Core Facility Director
Unit 3060
COR2E & Department of Chemistry
55 N Eagleville Road
University of Connecticut
Storrs, CT 06279

email: fko24003_at_uconn.edu<mailto:fko24003_at_uconn.edu>
(alias): luke.fulton_at_uconn.edu<mailto:luke.fulton_at_uconn.edu>
mobile: (603) 953-5275
Office: (860) 486-4069



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1247): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1247__;!!PvDODwlR4mBZyAb0!TYgui3kwFozpyRNfp64uG9G9Hmq1Ia=
0vCyU-siOelSTnX2P4RzuLwrkV9k4olDiNP3ilaLoenGTrmQ40nRw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106101163=
/7559972__;!!PvDODwlR4mBZyAb0!TYgui3kwFozpyRNfp64uG9G9Hmq1Ia0vCyU-siOelSTnX=
2P4RzuLwrkV9k4olDiNP3ilaLoenGTrhJulR5Y$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!TYgui3kwFozpyRNfp64uG9G9Hmq1Ia0vCy=
U-siOelSTnX2P4RzuLwrkV9k4olDiNP3ilaLoenGTrA-3yd5c$
Mute #troubleshooting:https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!TYgui3kwFozpyRNfp64u=
G9G9Hmq1Ia0vCyU-siOelSTnX2P4RzuLwrkV9k4olDiNP3ilaLoenGTrhSFl3J0$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue May 14 2024 - 12:26:49 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 01 2024 - 13:37:48 MST