From: Bob Berno <bberno_at_mcmaster.ca>
Date: Tue, 14 May 2024 09:52:40 -0400

Hello Muhammad,

 From the email thread, it looks like Issue 2 has been resolved.  You
just need to tease the signal of the smallest peaks out of the noise,
either by recollecting the data with better S/N or perhaps by some
processing tricks.

Regarding Issue 1:
You did not give any details about TopSpin version or operating system,
but I think I may have an idea why you need to keep running ATMM.  You
say that you run ATMM and "save" the values.  But do you know for sure
that the file containing the tune/match settings for 29Si have indeed
been saved?  I don't remember where those files get stored, but maybe
when it tries to save those settings, you don't have permission to
create and save a new file in that folder.  All other existing nuclei
work fine because they already exist.

Does that make sense?

Bob.

On 2024-05-14 3:26 a.m., Muhammad Adil Raees via groups.io wrote:
>
>
> Hello everyone,
> We recently conducted a 29Si experiment on our system and encountered
> two issues. We would appreciate your advice on these matters.
> Issue 1: I was able to tune and match the probe with ATMM as shown in
> the attached picture. I stored the values by clicking on the close and
> store values box. However, when I tried to perform ATMA, I was
> directed to optimize from ATMM again. A message box popped up on the
> screen, as attached in this email. I am wondering how to store the
> wobble curve so that ATMA can adjust it itself.
> Issue 2: The researcher expected a heptate from 29Si at around 188ppm,
> but I only detected a pentate. The signal is clear, but the outermost
> signals didn't appear. I am wondering what parameters I should focus
> on to improve the result. Do you think an inaccurate 90-degree pulse
> could be the cause?
>
> Kind regards,
>
> --
> Muhammad Adil Raees, Ph.D.
> Manager NMR Facility
> The Central Laboratory
> Syed Babar Ali School of Sciences and Engineering
> Lahore University of Management Sciences
> Opposite Sector-U, D.H.A. 54792, Lahore
> +92-330-2866755
> adil.raees_at_lums.edu.pk
>

--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
---------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.

McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One Spoon"
wampum agreement.


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1242): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1242__;!!PvDODwlR4mBZyAb0!UvgPFjqip3nEauqhBPkSleSiIqOwUI=
vL7wf2c7viY2st04ro_LImpdOdp0vWqvlF1pAi39YihXQT7y4M$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106091125=
/7559972__;!!PvDODwlR4mBZyAb0!UvgPFjqip3nEauqhBPkSleSiIqOwUIvL7wf2c7viY2st0=
4ro_LImpdOdp0vWqvlF1pAi39YihVIehM0a$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!UvgPFjqip3nEauqhBPkSleSiIqOwUIvL7wf2c7=
viY2st04ro_LImpdOdp0vWqvlF1pAi39YihV0NZdfB$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue May 14 2024 - 07:00:03 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 01 2024 - 13:37:48 MST