Re: [AMMRL] Topspin3.7 issues

From: <gregory.wylie_at_chem.tamu.edu>
Date: Mon, 13 May 2024 11:59:07 -0700

Hello All

So we did get Topspin 3.7 running on a new G4Z5 today with some of the issues
above.

We first had issues a few weeks ago with this system because we tried to put
3.7 on our older workstation (HP XW4600) as a test before moving to the new
workstation. Our firmware for the SampleCase was out of date so we could not
move forward. Bruker got us the firmware and the older workstation took 3.7
and we let it run for several weeks.

The new workstation had no issues putting on 3.7, we did do an nmr restore
before we did the cf. But as mentioned above cf failed with now spectrometer
on card 2. Rebooted IPSO no luck, restarted the computer no luck. Made sure
tftp and dhcpd were both running and they were. Reinstalled diskless and
rebooted IPSO again and cf completed! The wiring was all wrong but I can
live with that, an easy fix.

Anyway thanks for the info in this tread!

Thanks again to this community.

Greg
Gregory P. Wylie, Ph.D
NMR Facility Manager
Texas A&M University
Department of Chemistry
P.O. Box 30012
College Station, TX  77843-3255
gpwylie_at_tamu.edu
979.458.0705 (voice)
979.845.4719 (fax)
706.206.0007 (cell)
http://nmr.tamu.edu


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1235): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1235__;!!PvDODwlR4mBZyAb0!Wwi8PnZAwfiVm_xYPG4O0pDrJ5SUSv=
oKj3UkwH77Q4RZObBvpv7Tib-R_f1YzDfaHHYeuWvYbCzGVW0spfB0iSXDWTGo$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105862048=
/7559972__;!!PvDODwlR4mBZyAb0!Wwi8PnZAwfiVm_xYPG4O0pDrJ5SUSvoKj3UkwH77Q4RZO=
bBvpv7Tib-R_f1YzDfaHHYeuWvYbCzGVW0spfB0iflmnUbx$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon May 13 2024 - 11:59:10 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 01 2024 - 13:37:48 MST