Re: [AMMRL] Help with Mysterious Signals

From: Blumenfeld, Alexander (blum_at_uidaho.edu) <"Blumenfeld,>
Date: Sun, 5 May 2024 03:51:35 +0000

Hi Andre,

The first suspects are 13C-19F fragments. 500+ Hz coupling is kind of large,
but I remember seeing coupling of around 450 Hz. The simple way to check it
is to run a quick 19F spectrum and look for 13C satellites with the same
splitting. I don't see in your list any spin ½ metal that can give
such splitting.

Good luck,

Alex
*********​​*****
Alex Blumenfeld
NMR Lab Manager
University of Idaho
Moscow, ID 83844
208-885-9235
________________________________
> From: main_at_ammrl.groups.io on behalf of Andre Simpson
> Sent: Saturday, May 4, 2024 8:56 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] Help with Mysterious Signals

Hi Ammrl,

Sorry to bother you all. I am hoping to draw on your extensive experience. =
We are doing a large project where we are using NMR alongside MS to discover
unexpected discharges into the environment.

We have these very weird large signals in 1H NMR. They are quite strong even
in the unconcentrated effluent. I just cannot work out what they may be.
They are exclusive to effluents from the electroplating industry and appear
in some (but not all) discharges.

I have attached, a sheet of what we know. I am hoping someone somewhere may
have seen something like this before and can point us the right direction ?

Thanks so so much for your time, Cheers, Andre

Professor Andre Simpson, PhD, FRSC

Director of the Environmental NMR Center, Founder of TRACES
Department of Chemistry
University of Toronto, Scarborough Campus
1265 Military Trail
Toronto, Canada
M1C 1A4

andre.simpson_at_utoronto.ca
https://www.utsc.utoronto.ca/labs/asimpson/


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1227): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1227__;!!PvDODwlR4mBZyAb0!R4tZXRKjVRuKnuLgHAV1OGSLCN_6IS=
-p5bTetOiM76WpUkM36bHST7m96_qKAHS58GCfxBSQAqXDnw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105914179=
/7559972__;!!PvDODwlR4mBZyAb0!R4tZXRKjVRuKnuLgHAV1OGSLCN_6IS-p5bTetOiM76WpU=
kM36bHST7m96_qKAHS58GCfxBQ4AQTZHQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Sun May 05 2024 - 06:39:58 MST

This archive was generated by hypermail 2.4.0 : Sat Jun 01 2024 - 13:37:48 MST