Thanks for the feedback on my question. As pointed out by multiple
people, I should have checked the tuning range of the probe (Prodigy
cryoprobe) first before trying to set up anything else.  That
information is actually readily available in edprobe under "coil
parameters".  As 6Li is close to deuterium it is excluded with that
probe, as it is with the iProbe on our 400MHz instrument.

When changing the current probe to our backup Smart Probe instead of
Prodigy, 6Li will appear in the list of nuclei edprosol, so if we down
the rod decide we need 6Li we can do it. But for now 7Li will just do fine.

Thanks for responding

Holger

On 2/11/2026 12:29 PM, Frank Holger Foersterling via groups.io wrote:

> Hello everybody,
> I am trying to run a new nucleus (6Li) for the first time on a Bruker
> Avance III HD, running Topspin 3.8. I notice that there is no entry
> for that nucleus in either the PROSOL (edprosol) or PeakPower table
> (edprobe). I tried adding an entry in the PeakPower list, but the
> nucleus did not show up as an option in the pulldown menu, and there
> seems to be no way to add another nucleus in thje PROSOL table.
> I was able to copy a CORETAB entry (17O) to 6Li in
> TS/conf/instr/spec/cortab, which is recognized in cortab, but it does
> not seem to add that nucleus as an option to the PROSOL table.
> It does show up in ednuc as a known nucleus.
> I was able to get entries for 7Li into both PeakPower and Prosol, but
> some references indicate that 6Li may give sharper lines sometimes, so
> I was going to give that a try also.
> Thanks for any advise
> Holger Foersterling
>


--
Frank H. Foersterling
Instrumentation Specialist
Chem-B87
Department of Chemistry
University of Wisconsin-Milwaukee
414-229-5896(Lab) 414-229-5035 (Teams)
holger_at_uwm.edu


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3287): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3287__;!!PvDODwlR4mBZyAb0!U175qeLjwd6BwVzwp5qdhKpmEKsDbl=
lzku_gQwvU_lZXxh94UrsfIIz7enXNNKOt8jpMVfzMr9p7Q1gTY_7iBsrH1g$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117761768=
/7559972__;!!PvDODwlR4mBZyAb0!U175qeLjwd6BwVzwp5qdhKpmEKsDbllzku_gQwvU_lZXx=
h94UrsfIIz7enXNNKOt8jpMVfzMr9p7Q1gTY_7ISR0JUg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Mar 11 2026 - 10:24:25 MST

This archive was generated by hypermail 2.4.0 : Wed Apr 01 2026 - 17:02:32 MST