[AMMRL] Signal Blanking on High Band

From: Shawn Wagner via groups.io <shawn.wagner=usc.edu_at_groups.io>
Date: Wed, 11 Mar 2026 16:48:59 +0000

Hi AMMRL Community,

Before I go hunting for the problem, I wanted to check if anyone has seen
this before. I have an unusual issue on a Varian 400MR (DirectDrive).

The signal on the high-band channel (channel 1, 1H/19F) blanks for ~0.125 s
every ~1.6 s. During the blanked period the receiver output goes completely
flat. This does not occur on channel 2.

When tuning the high-band channel you can see the same behavior. If the
blanking occurs during the tune sweep, the signal disappears and only noise is
visible.

Things I have checked so far:

  *
Swapped the XMTR and ATTEN/SW boards $B"*(B no change
  *
The problem occurs even with no transmit pulse
  *
Verified the probe using a network analyzer
  *
The issue is only on channel 1

One additional observation: if I turn the transmitter pulse off and set the
receiver gain to 0, the blanked portion of the FID (~0.125 s) actually shows
more noise than when it is acquiring normally.

Has anyone seen this type of periodic receiver blanking on a 400MR? If someone
recognizes the symptom and can save me from rooting through the console internals,
I would greatly appreciate it.

Thanks,

Shawn



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3288): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3288__;!!PvDODwlR4mBZyAb0!S0fTaHBraZxBn1uKdcCRiyftaCyZgH=
7mnR_NobtyH5tLbv5nlIBZw67GCjJGDs9-bqFvm1i0E_9DRBWJumC80msJO6Qk$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118264535=
/7559972__;!!PvDODwlR4mBZyAb0!S0fTaHBraZxBn1uKdcCRiyftaCyZgH7mnR_NobtyH5tLb=
v5nlIBZw67GCjJGDs9-bqFvm1i0E_9DRBWJumC80kUnNC1c$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Mar 11 2026 - 10:24:33 MST

This archive was generated by hypermail 2.4.0 : Wed Apr 01 2026 - 17:02:32 MST