Hi,
We have recently upgraded to TopSpin 3.7.0 when we replaced an aging computer
with a new linux system run Alma9. We are seeing the same issue with the
SampleJet
When IconNMR is used and a user exits cleanly, commands from the TopSpin command
line to insert samples are not recognized. The sx <holder_number> doesn’t work
and we get a communications error. We can still see and communicate with the SampleJet
using the web interface and a restart of the application from the web interface
resolves the issue.
Best,
Francis
_________________________________________
Francis Peterson, Professor
Department of Biochemistry
Medical College of Wisconsin
8701 Watertown Plank Road
Milwaukee, WI 53226
(414)-955-5777(office)/8081(lab)/456-6510(fax)
mailto:fpeterso_at_mcw.edu
__________________________________________
On Mar 27, 2025, at 8:40 PM, Anklin, Clemens via groups.io wrote:
Hi Cathy
the topshim issue is a bug that has recently popped up. The only way to fix
it is to go into the file that controls the defaults and change the parameter
that controls the dimension.
The file is in C:\users\username]\.topspin1\prop or /home/username/.topspin1/prop
and is called topshim.prop . In there is a line called Dimension. False is 1D and
true is 3D. Change it to false and store it. Then it will the topshim window will
pop up with 1D as default.
As for the samplejet do you mean that a command like sx <holder number> does not work.
Do you get an error when you give this command.
Best regards
Clemens
On 3/27/2025 5:08 PM, Cathy Clewett via groups.io wrote:
Hi all,
We just installed Topspin 3.7.0 with our upgrade to Alma9 on our Avance III
400 MHz nanobay. Most things are working pretty well, but we've run into
two problem that are stumping us.
When we open Topshim GUI, the default has it marked as a 3D shim. We want
to change that to a 1D shim, but when we go to the service panel and ask it
to save it, it doesn't. We've tried reloading the defaults, and saving those,
but it doesn't make the changes. Command line topshim is working fine. In fact
topshim is working fine, we just have to change the 1D radio button (and I
doubt our standard user will do that).
The other problem is that If a user starts Icon and exits normally, we can
no longer talk to our SampleJet through the TopSpin command line interface.
If we go through a complete error recovery, we can talk to the SampleJet until
we open Icon. Using Icon seems to always be able to talk to the Sample Jet.
Things we have tried
*
reboot of the spectrometer
*
Once everything was on, cf
*
Reinstalling previous firmware
Nothing has worked so far. I'm just about ready to reinstall Topspin 3.7.0,
thinking that I tested too many things as a user instead of nmrsu. Before
I do that, I'd love to hear if anybody has any suggestions. (Uncharacteristically,
Rover isn't getting back to me.)
Thanks all,
Cathy
Catherine F. M. Clewett, Ph.D.
Asst. Director Magnetic Resonance
Chemistry Instrumentation Center
University of Wisconsin Madison, Department of Chemistry
1101 University Ave Room 2201c, Madison, WI 53706
clewett_at_wisc.edu,
(608)262-8196,
https://pbcic.chem.wisc.edu
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2177): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2177__;!!PvDODwlR4mBZyAb0!X42tMX_0unnNdM-0_Jpgg5Q6NsmYn-=
fxwRIk1h7BPuOnSG4OnGe5b-7LQZOKv3aun1FCSsAxiRyTEUqbhYO5ydbukLdB$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111947639=
/7559972__;!!PvDODwlR4mBZyAb0!X42tMX_0unnNdM-0_Jpgg5Q6NsmYn-fxwRIk1h7BPuOnS=
G4OnGe5b-7LQZOKv3aun1FCSsAxiRyTEUqbhYO5yWsOHsy5$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!X42tMX_0unnNdM-0_=
Received on Fri Mar 28 2025 - 04:51:44 MST