[AMMRL] Can't save defaults on Topshim 2.4.0 with Topspin 3.7.0 and SampleJet gets lost for manual control after being used with Icon

From: Cathy Clewett via groups.io <clewett=wisc.edu_at_groups.io>
Date: Thu, 27 Mar 2025 21:08:13 +0000

Hi all,

We just installed Topspin 3.7.0 with our upgrade to Alma9 on our Avance III
400 MHz nanobay. Most things are working pretty well, but we've run into
two problem that are stumping us.

When we open Topshim GUI, the default has it marked as a 3D shim. We want
to change that to a 1D shim, but when we go to the service panel and ask it
to save it, it doesn't. We've tried reloading the defaults, and saving those,
but it doesn't make the changes. Command line topshim is working fine.
In fact topshim is working fine, we just have to change the 1D radio button
(and I doubt our standard user will do that).

The other problem is that If a user starts Icon and exits normally, we can
no longer talk to our SampleJet through the TopSpin command line interface.
If we go through a complete error recovery, we can talk to the SampleJet
until we open Icon. Using Icon seems to always be able to talk to the Sample
Jet. Things we have tried

  *
reboot of the spectrometer
  *
Once everything was on, cf
  *
Reinstalling previous firmware

Nothing has worked so far. I'm just about ready to reinstall Topspin 3.7.0,
thinking that I tested too many things as a user instead of nmrsu. Before I
do that, I'd love to hear if anybody has any suggestions. (Uncharacteristically,
Rover isn't getting back to me.)

Thanks all,

Cathy

Catherine F. M. Clewett, Ph.D.
Asst. Director Magnetic Resonance
Chemistry Instrumentation Center
University of Wisconsin Madison, Department of Chemistry
1101 University Ave Room 2201c, Madison, WI 53706

clewett_at_wisc.edu<mailto:clewett_at_wisc.edu>,
(608)262-8196,
https://pbcic.chem.wisc.edu




-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2175): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2175__;!!PvDODwlR4mBZyAb0!UtltmoW9rDK9fIkBZwQ09peM2fyaNi=
2KBuLtaEYItpWHPHINy00pzmPwY9SKBa7jFaJKmoa2zwN-XKVm770Opl08j6YU$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111947639=
/7559972__;!!PvDODwlR4mBZyAb0!UtltmoW9rDK9fIkBZwQ09peM2fyaNi2KBuLtaEYItpWHP=
HINy00pzmPwY9SKBa7jFaJKmoa2zwN-XKVm770OpkFEIKO-$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Mar 27 2025 - 18:25:08 MST

This archive was generated by hypermail 2.4.0 : Tue Apr 01 2025 - 14:21:16 MST